The x-ray sensitivity of amorphous selenium for mammography

被引:66
作者
Stone, MF
Zhao, W
Jacak, BV
O'Connor, P
Yu, B
Rehak, P
机构
[1] SUNY Stony Brook, Hlth Sci Ctr 120, Dept Radiol, Stony Brook, NY 11794 USA
[2] SUNY Stony Brook, Dept Phys & Astron, Stony Brook, NY 11794 USA
[3] Brookhaven Natl Lab, Instrumentat Div, Upton, NY 11973 USA
关键词
amorphous selenium; x-ray sensitivity; mammography;
D O I
10.1118/1.1449874
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A study of the x-ray sensitivity of amorphous selenium (a-Se) for digital mammography has been performed. A uniform layer of a-Se was deposited on a glass substrate with electrodes on both surfaces. The deposition procedure was identical to that used for a-Se flat-panel detectors. A high voltage was applied to the top surface of the a-Se layer in order to establish an electric field E-Se. Then the sample was exposed to x rays with 27 kVp spectra generated from an x-ray tube with a molybdenum (Mo) target. The mean x-ray energy of the spectrum used was similar to16.6 keV. The x-ray current generated by the a-Se layer was measured as a function of E-Se. From the current measurement and the estimation of total x-ray energy absorbed in the a-Se, the energy required to create one electron-hole pair (EHP), W, was determined as a function of E-Se. It was found that at the most commonly used E-Se of 10 V/mum, W was measured as 64 eV. This is considerably higher than the widely accepted typical value of W = 50 eV measured at higher x-ray photon energies (e.g., 50 keV). The dependence of W as a function of E-Se can be best fitted using the empirical expression of E-Se(-y). This relationship is consistent with the results obtained at higher x-ray energies. This article provides an accurate measurement of x-ray sensitivity of a-Se at mammographic energies independent of detector operation, such as the most recently developed flat-panel detectors. The results will be a useful toot for investigation and optimization of a-Se-based x-ray imaging detectors, such as determination of pixel fill-factor and optimal E-Se during operation. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:319 / 324
页数:6
相关论文
共 30 条
[21]   EFFECT OF FINITE PHOSPHOR THICKNESS ON DETECTIVE QUANTUM EFFICIENCY [J].
NISHIKAWA, RM ;
YAFFE, MJ ;
HOLMES, RB .
MEDICAL PHYSICS, 1989, 16 (05) :773-780
[22]   Direct conversion detector for digital mammography [J].
Polischuk, B ;
Rougeot, H ;
Wong, K ;
Debrie, A ;
Poliquin, E ;
Hansroul, M ;
Martin, JP ;
Truong, TT ;
Choquette, M ;
Laperrière, L ;
Shukri, Z .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :417-425
[23]  
Rowlands J.A., 2000, Handbook of Medical Imaging, V1, P223, DOI [https://doi.org/10.1117/3.832716, DOI 10.1117/3.832716.CH4, DOI 10.1117/3.832716]
[24]   X-RAY-IMAGING USING AMORPHOUS SELENIUM - A PHOTOINDUCED DISCHARGE READOUT METHOD FOR DIGITAL MAMMOGRAPHY [J].
ROWLANDS, JA ;
HUNTER, DM ;
ARAJ, N .
MEDICAL PHYSICS, 1991, 18 (03) :421-431
[25]   X-RAY-IMAGING USING AMORPHOUS SELENIUM - DETERMINATION OF X-RAY-SENSITIVITY BY PULSE-HEIGHT SPECTROSCOPY [J].
ROWLANDS, JA ;
DECRESCENZO, G ;
ARAJ, N .
MEDICAL PHYSICS, 1992, 19 (04) :1065-1069
[26]   Digital X-ray systems based on amorphous selenium [J].
Rowlands, JA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (02) :409-411
[27]   X-ray imaging using lead iodide as a semiconductor detector [J].
Street, RA ;
Rahn, JT ;
Ready, SE ;
Shah, K ;
Bennett, P ;
Dmitriyev, Y ;
Mei, P ;
Lu, JP ;
Apte, RB ;
Ho, J ;
van Schuylenbergh, K ;
Lemmi, F ;
Boyce, JB ;
Nylen, P .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :36-47
[28]   Development and evaluation of a large-area selenium-based flat panel detector for real-time radiography and fluoroscopy [J].
Tsukamoto, A ;
Yamada, S ;
Tomisaki, T ;
Tanaka, M ;
Sakaguchi, T ;
Asahina, H ;
Suzuki, K ;
Ikeda, M .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :14-23
[29]  
YAFFE MJ, 1999, CATEGORICAL COURSE D, P229
[30]   X-RAY-IMAGING USING AMORPHOUS SELENIUM - FEASIBILITY OF A FLAT-PANEL SELF-SCANNED DETECTOR FOR DIGITAL RADIOLOGY [J].
ZHAO, W ;
ROWLANDS, JA .
MEDICAL PHYSICS, 1995, 22 (10) :1595-1604