Image Fusion in Dual Energy Computed Tomography Effect on Contrast Enhancement, Signal-to-Noise Ratio and Image Quality in Computed Tomography Angiography

被引:132
作者
Behrendt, Florian F. [1 ]
Schmidt, Bernhard [2 ]
Plumhans, Cedric [1 ]
Keil, Sebastian [1 ]
Woodruff, Seth G. [3 ]
Ackermann, Diana [4 ]
Muehlenbruch, Georg [1 ]
Flohr, Thomas [2 ]
Guenther, Rolf W. [1 ]
Mahnken, Andreas H. [1 ]
机构
[1] Univ Hosp RWTH Aachen, Dept Diagnost Radiol, D-52057 Aachen, Germany
[2] Siemens Healthcare Computed Tomog, Forchheim, Germany
[3] Univ Hosp RWTH Aachen, Dept Obstet & Gynecol, D-52057 Aachen, Germany
[4] Univ Hosp RWTH Aachen, Inst Med Stat, D-52057 Aachen, Germany
关键词
dual energy CT (DECT); image fusion; contrast enhancement; signal-to-noise ratio (SNR); image quality; LOW TUBE VOLTAGE; CT; DIFFERENTIATION; EXPERIENCE; STONES;
D O I
10.1097/RLI.0b013e31818c3d4b
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this study was to evaluate the influence of different weighting factors on contrast enhancement, signal-to-noise ratio (SNR), and image quality in image fusion in:dual energy computed tomography (DECT) angiography. Material and Methods: Fifteen patients underwent a CT angiography of the aorta with a SOMATOM Definition Dual Source CT (DSCT; Siemens, Forchheim, Germany) in dual energy mode (DECT) (tube voltage: 80 and 140 kVp; tube current: 297 eff. mA. and 70 eff. mA; collimation, 14 x 1.2 min). Raw data were reconstructed using a soft convolution kernel (D30f). Fused images were calculated using a spectrum of weighting factors (0.0, 0.1, 0.3, 0.5, 0.7, 0.9, and 1.0) generating different ratios between the 80- and 140-kVp images (eg, factor 0.5 corresponds to 50% image information from the 140- and the 80-kVp image). Both CT values and SNR were measured in the descending aorta (levels of celiac trunk, renal arteries, and aortic bifurcation), in the right and left common iliac artery and in paraaortal fat. Image quality was evaluated using a 5-point grading scale. Results were compared using paired t-tests and nonparametric paired Wilcoxon tests. Results: Statistically significant increases in mean CT values were seen in vessels when increasing weighting factors were used (all P <= 0.001). For example, mean CT values derived from the aorta at the level of the celiac trunk were 273.8 +/- 25.8 Hounsfield units (HU), 304.0 +/- 24.3 HU, 361.4 +/- 22.5 HU, 418.3 +/- 25.8 HU, 477.8 +/- 32.2 HU, 536.2 +/- 41.2 HU, 564.6 +/- 45.3 HU, when the weighting factors 0.0, 0.1, 0.3, 0.5, 0.7, 0.9, and 1.0 were used. The highest SNR values were found in vessels when the weighting factor 0.5 was used. The highest SNR values of the paraaortal fat were obtained for the weighting factors 0.3 and 0.5. Visual image assessment for image quality showed the highest score for the data reconstructed using the weighting factor 0.5. Conclusion: Different weighting factors used to create fused images in DECT cause statistically significant differences in CT value, SNR, and image quality. Best results were obtained using the weighting factor 0.5, which we recommend for image fusion in DECT angiography.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 10 条
[1]   QUANTITATIVE THEORY OF HOUNSFIELD UNIT AND ITS APPLICATION TO DUAL ENERGY SCANNING [J].
BROOKS, RA .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (04) :487-493
[2]   Dual energy CT characterization of urinary calculi: Initial in vitro and clinical experience [J].
Graser, Anno ;
Johnson, Thorsten R. C. ;
Bader, Markus ;
Staehler, Michael ;
Haseke, Nicolas ;
Nikolaou, Konstantin ;
Reiser, Maximilian F. ;
Stief, Christian G. ;
Becker, Christoph R. .
INVESTIGATIVE RADIOLOGY, 2008, 43 (02) :112-119
[3]   Characterization of human renal stones with MDCT: Advantage of dual energy and limitations due to respiratory motion [J].
Grosjean, Romain ;
Sauer, Benoit ;
Guerra, Rui Matias ;
Daudon, Michel ;
Blum, Alain ;
Felblinger, Jacques ;
Hubert, Jacques .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (03) :720-728
[4]   Material differentiation by dual energy CT: initial experience [J].
Johnson, Thorsten R. C. ;
Krauss, Bernhard ;
Sedlmair, Martin ;
Grasruck, Michael ;
Bruder, Herbert ;
Morhard, Dominik ;
Fink, Christian ;
Weckbach, Sabine ;
Lenhard, Miriam ;
Schmidt, Bernhard ;
Flohr, Thomas ;
Reiser, Maximilian F. ;
Becker, Christoph R. .
EUROPEAN RADIOLOGY, 2007, 17 (06) :1510-1517
[5]   Low tube voltage improves computed tomography imaging of delayed myocardial contrast enhancement in an experimental acute myocardial infarction model [J].
Mahnken, Andreas H. ;
Bruners, Philipp ;
Muehlenbruch, Georg ;
Emmerich, Melanie ;
Hohl, Christian ;
Guenther, Rolf W. ;
Wildberger, Joachim E. .
INVESTIGATIVE RADIOLOGY, 2007, 42 (02) :123-129
[6]   Abdominal CT with low tube voltage: Preliminary observations about radiation dose, contrast enhancement, image quality, and noise [J].
Nakayama, Y ;
Awai, K ;
Funama, Y ;
Hatemura, M ;
Imuta, M ;
Nakaura, T ;
Ryu, D ;
Morishita, S ;
Sultana, S ;
Sato, N ;
Yamashita, Y .
RADIOLOGY, 2005, 237 (03) :945-951
[7]   Noninvasive differentiation of uric acid versus non-uric acid kidney stones using dual-energy CT [J].
Primak, Andrew N. ;
Fletcher, Joel G. ;
Vrtiska, Terri J. ;
Dzyubak, Oleksandr P. ;
Lieske, John C. ;
Jackson, Molly E. ;
Williams, James C., Jr. ;
McCollough, Cynthia H. .
ACADEMIC RADIOLOGY, 2007, 14 (12) :1441-1447
[8]   Multislice CT angiography [J].
Prokop, M .
EUROPEAN JOURNAL OF RADIOLOGY, 2000, 36 (02) :86-96
[9]   Dual-energy contrast-enhanced computed tomography for the detection of urinary stone disease [J].
Scheffel, Hans ;
Stolzmann, Paul ;
Frauenfelder, Thomas ;
Schertler, Thomas ;
Desbiolles, Lotus ;
Leschka, Sebastian ;
Marincek, Borut ;
Alkadhi, Hatem .
INVESTIGATIVE RADIOLOGY, 2007, 42 (12) :823-829
[10]   Effect of x-ray tube parameters, iodine concentration, and patient size on image quality in pulmonary computed tomography angiography: A chest-phantom-study [J].
Szucs-Farkas, Zsolt ;
Verdun, Francis R. ;
von Allmen, Gabriel ;
Mini, Roberto L. ;
Vock, Peter .
INVESTIGATIVE RADIOLOGY, 2008, 43 (06) :374-381