The reproducibility of organ position using active breathing control (ABC) during liver radiotherapy

被引:236
作者
Dawson, LA
Brock, KK
Kazanjian, S
Fitch, D
McGinn, CJ
Lawrence, TS
Ten Haken, RK
Balter, J
机构
[1] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Radiol, Ann Arbor, MI 48109 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2001年 / 51卷 / 05期
关键词
active breathing control; intrahepatic cancer; conformal radiation; interfraction reproducibility; intrafraction reproducibility;
D O I
10.1016/S0360-3016(01)02653-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the intrafraction and interfraction reproducibility of liver immobilization using active breathing control (ABC). Methods and Materials: Patients with unresectable intrahepatic tumors who could comfortably hold their breath for at least 20 s were treated with focal liver radiation using ABC for liver immobilization. Fluoroscopy was used to measure any potential motion during ABC breath holds. Preceding each radiotherapy fraction, with the patient setup in the nominal treatment position using ABC, orthogonal radiographs were taken using room-mounted diagnostic X-ray tubes and a digital imager. The radiographs were compared to reference images using a 2D alignment tool. The treatment table was moved to produce acceptable setup, and repeat orthogonal verification images were obtained. The positions of the diaphragm and the liver (assessed by localization of implanted radiopaque intra-arterial microcoils) relative to the skeleton were subsequently analyzed. The intrafraction reproducibility (from repeat radiographs obtained within the time period of one fraction before treatment) and interfraction reproducibility (from comparisons of the first radiograph for each treatment with a reference radiograph) of the diaphragm and the hepatic microcoil positions relative to the skeleton with repeat breath holds using ABC were then measured. Caudal-cranial (CC), anterior-posterior (AP), and medial-lateral (ML) reproducibility of the hepatic microcoils relative to the skeleton were also determined from three-dimensional alignment of repeat CT scans obtained in the treatment position. Results: A total of 262 fractions of radiation were delivered using ABC breath holds in 8 patients. No motion of the diaphragm or hepatic microcoils was observed on fluoroscopy during ABC breath holds. From analyses of 158 sets of positioning radiographs, the average intrafraction CC reproducibility (sigma) of the diaphragm and hepatic microcoil position relative to the skeleton using ABC repeat breath holds was 2.5 mm (range 1.8-3.7 min) and 2.3 mm (range 1.2-3.7 mm) respectively. However, based on 262 sets of positioning radiographs, the average interfraction CC reproducibility (sigma) of the diaphragm and hepatic microcoils was 4.4 mm (range 3.0-6.1 mm) and 4.3 mm (range 3.1-5.7 mm.), indicating a change of diaphragm and microcoil position relative to the skeleton over the course of treatment with repeat breath holds at the same phase of the respiratory cycle. The average population absolute intrafraction CC offset in diaphragm and microcoil position relative to skeleton was 2.4 mm and 2.1 mm respectively; the average absolute interfraction CC offset was 5.2 mm. Analyses of repeat CT scans demonstrated that the average intrafraction excursion of the hepatic microcoils relative to the skeleton in the CC, AP, and NIL directions was 1.9 mm, 0.6 mm, and 0.6 mm respectively and the average interfraction CC, AP, and ML excursion of the hepatic microcoils was 6.6 mm, 3.2 mm, and 3.3 min respectively. Conclusion: Radiotherapy using ABC for patients with intrahepatic cancer is feasible, with good intrafraction reproducibility of liver position using ABC. However, the interfraction reproducibility of organ position with ABC suggests the need for daily on-line imaging and repositioning if treatment margins smaller than those required for free breathing are a goal. (C) 2001 Elsevier Science Inc.
引用
收藏
页码:1410 / 1421
页数:12
相关论文
共 34 条
  • [21] Respiratory gated irradiation system for heavy-ion radiotherapy
    Minohara, S
    Kanai, T
    Endo, M
    Noda, K
    Kanazawa, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (04): : 1097 - 1103
  • [22] IRRADIATION SYNCHRONIZED WITH RESPIRATION GATE
    OHARA, K
    OKUMURA, T
    AKISADA, M
    INADA, T
    MORI, T
    YOKOTA, H
    CALAGUAS, MJB
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (04): : 853 - 857
  • [23] OSAKA Y, 1997, 12 INT C US COMP RAD, P176
  • [24] IMPACT OF TUMOR-CONTROL ON SURVIVAL IN CARCINOMA OF THE LUNG TREATED WITH IRRADIATION
    PEREZ, CA
    BAUER, M
    EDELSTEIN, S
    GILLESPIE, BW
    BIRCH, R
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1986, 12 (04): : 539 - 547
  • [25] PREDICTIVE RESPIRATORY GATING - A NEW METHOD TO REDUCE MOTION ARTIFACTS ON CT SCANS
    RITCHIE, CJ
    HSIEH, J
    GARD, MF
    GODWIN, JD
    KIM, YM
    CRAWFORD, CR
    [J]. RADIOLOGY, 1994, 190 (03) : 847 - 852
  • [26] The deep inspiration breath-hold technique in the treatment of inoperable non-small-cell lung cancer
    Rosenzweig, KE
    Hanley, J
    Mah, D
    Mageras, G
    Hunt, M
    Toner, S
    Burman, C
    Ling, CC
    Mychalczak, B
    Fuks, Z
    Leibel, SA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (01): : 81 - 87
  • [27] Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor
    Shirato, H
    Shimizu, S
    Kitamura, K
    Nishioka, T
    Kagei, K
    Hashimoto, S
    Aoyama, H
    Kunieda, T
    Shinohara, N
    Dosaka-Akita, H
    Miyasaka, K
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (02): : 435 - 442
  • [28] CRANIO-CAUDAL MOVEMENTS OF THE LIVER, PANCREAS AND KIDNEYS IN RESPIRATION
    SURAMO, I
    PAIVANSALO, M
    MYLLYLA, V
    [J]. ACTA RADIOLOGICA-DIAGNOSIS, 1984, 25 (02): : 129 - 131
  • [29] Potential benefits of eliminating planning target volume expansions for patient breathing in the treatment of liver tumors
    TenHaken, RK
    Balter, JM
    Marsh, LH
    Robertson, JM
    Lawrence, TS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03): : 613 - 617
  • [30] Intrafractional tumor position stability during computed tomography (CT)-guided frameless stereotactic radiation therapy for lung or liver cancers with a fusion of CT and linear accelerator (FOCAL) unit
    Uematsu, M
    Shioda, A
    Suda, A
    Tahara, K
    Kojima, T
    Hama, Y
    Kong, M
    Wong, JR
    Fukui, T
    Kusano, S
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (02): : 443 - 448