Laser Ablation Coupled to a Flowing Atmospheric Pressure Afterglow for Ambient Mass Spectral Imaging

被引:92
作者
Shelley, Jacob T. [1 ]
Ray, Steven J. [1 ]
Hieftje, Gary M. [1 ]
机构
[1] Indiana Univ, Dept Chem, Bloomington, IN 47405 USA
关键词
D O I
10.1021/ac801594u
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A plasma-based ambient desorption/ionization mass spectrometry (ADI-MS) source was used to perform molecular mass spectral imaging. A small amount of sample material was ablated by focusing 266 nm laser light onto a spot. The resulting aerosol was transferred by a nitrogen stream to the flowing afterglow of a helium atmospheric pressure glow discharge ionization source; the ionized sample material was analyzed by a Leco Unique time-of-flight mass spectrometer. Two-dimensional mass spectral images were generated by scanning the laser beam across a sample surface. The total analysis time for a 6 mm(2) surface, which is limited by the washout of the ablation chamber, was less than 30 min. With this technique, a spatial resolution of similar to 20 pin has been achieved. Additionally, the laser ablation configuration was used to obtain depth information of over 2 min with a resolution of similar to 40 pm. The combination of laser ablation with the flowing atmospheric pressure afterglow source was used to analyze several sample surfaces for a wide variety of analytes and with high sensitivity (LOD of 5 fmol for caffeine).
引用
收藏
页码:8308 / 8313
页数:6
相关论文
共 21 条
[1]   Atmospheric pressure chemical ionization source. 2. Desorption-ionization for the direct analysis of solid compounds [J].
Andrade, Francisco J. ;
Shelley, Jacob T. ;
Wetzel, William C. ;
Webb, Michael R. ;
Gamez, Gerardo ;
Ray, Steven J. ;
Hieftje, Gary M. .
ANALYTICAL CHEMISTRY, 2008, 80 (08) :2654-2663
[2]   Atmospheric pressure chemical ionization source. 1. Ionization of compounds in the gas phase [J].
Andrade, Francisco J. ;
Shelley, Jacob T. ;
Wetzel, William C. ;
Webb, Michael R. ;
Gamez, Gerardo ;
Ray, Steven J. ;
Hieftje, Gary M. .
ANALYTICAL CHEMISTRY, 2008, 80 (08) :2646-2653
[3]   Use of a novel array detector for the direct analysis of solid samples by laser ablation inductively coupled plasma sector-field mass spectrometry [J].
Barnes, JH ;
Schilling, GD ;
Hieftje, GM ;
Sperline, RP ;
Denton, MB ;
Barinaga, CJ ;
Koppenaal, DW .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2004, 15 (06) :769-776
[4]   Depth profile analysis of various titanium based coatings on steel and tungsten carbide using laser ablation inductively coupled plasma -: "time of flight" mass spectrometry [J].
Bleiner, D ;
Plotnikov, A ;
Vogt, C ;
Wetzig, K ;
Günther, D .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 368 (2-3) :221-226
[5]   Molecular imaging of biological samples: Localization of peptides and proteins using MALDI-TOF MS [J].
Caprioli, RM ;
Farmer, TB ;
Gile, J .
ANALYTICAL CHEMISTRY, 1997, 69 (23) :4751-4760
[6]   Versatile new ion source for the analysis of materials in open air under ambient conditions [J].
Cody, RB ;
Laramée, JA ;
Durst, HD .
ANALYTICAL CHEMISTRY, 2005, 77 (08) :2297-2302
[7]   Ambient mass spectrometry [J].
Cooks, RG ;
Ouyang, Z ;
Takats, Z ;
Wiseman, JM .
SCIENCE, 2006, 311 (5767) :1566-1570
[8]   Simulation of atmospheric transport and droplet-thin film collisions in desorption electrospray ionization [J].
Costa, Anthony B. ;
Cooks, R. Graham .
CHEMICAL COMMUNICATIONS, 2007, (38) :3915-3917
[9]   Forensic analysis of inks by imaging desorption electrospray ionization (DESI) mass spectrometry [J].
Ifa, D. R. ;
Gumaelius, L. M. ;
Eberlin, L. S. ;
Manicke, N. E. ;
Cooks, R. G. .
ANALYST, 2007, 132 (05) :461-467
[10]  
KLINKERT I, 2007, REV SCI INSTRUM, V78