Preliminary results with a new vascular basket occluder in swine

被引:3
作者
Kónya, A [1 ]
Wright, KC [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Diagnost Radiol, Houston, TX 77030 USA
关键词
arteries; therapeutic blockade; embolization;
D O I
10.1016/S1051-0443(99)70190-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To experimentally evaluate a new self-anchoring vascular occlusion device suitable for use in large, high-flow vessels. MATERIALS AND METHODS: A basket-shaped occluding device consisting of a nitinol frame (with or without polyester threads) covered with Dacron was evaluated in the common iliac arteries and abdominal aortae of normal pigs, One occluder was placed in each recipient vessel. Angiography was performed before and after placement of each occluder. RESULTS: Precise placement of the device was achieved in all cases. Mechanically, all but one occluder worked dependably. Angiographically, devices without polyester threads produced only partial occlusion, Devices with threads produced complete occlusion in all but one abdominal aorta, in which partial occlusion occurred because the occluder opened asymmetrically. Complete iliac arterial occlusion occurred within 2-8 minutes of placement and aortic occlusion was achieved within 8-20 minutes. At necropsy, all devices were found to be securely anchored against the vascular wall, CONCLUSION: Preliminary results indicate that this vascular basket occluder containing polyester threads is easy to place precisely, produces rapid embolization, and demonstrates good self-anchoring ability in a high-sow arterial model.
引用
收藏
页码:1043 / 1049
页数:7
相关论文
共 19 条
[1]   BIOCOMPATIBILITY OF NITINOL ALLOY AS AN IMPLANT MATERIAL [J].
CASTLEMAN, LS ;
MOTZKIN, SM ;
ALICANDRI, FP ;
BONAWIT, VL ;
JOHNSON, AA .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1976, 10 (05) :695-731
[2]   Endovascular occlusion with a new mechanical detachable coil [J].
Coley, SC ;
Jackson, JE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 171 (04) :1075-1079
[3]   PULMONARY ARTERIOVENOUS-MALFORMATIONS - RESULTS OF TREATMENT WITH COIL EMBOLIZATION IN 53 PATIENTS [J].
DUTTON, JAE ;
JACKSON, JE ;
HUGHES, JMB ;
WHYTE, MKB ;
PETERS, AM ;
USSOV, W ;
ALLISON, DJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (05) :1119-1125
[4]   MECHANICAL DEVICES FOR ARTERIAL-OCCLUSION [J].
GIANTURCO, C ;
ANDERSON, JH ;
WALLACE, S .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1975, 124 (03) :428-435
[5]   Transcatheter patent ductus arteriosus closure in an infant using the Gianturco-Grifka Vascular Occlusion Device [J].
Grifka, RG ;
Vincent, JA ;
Nihill, MR ;
Ing, FF ;
Mullins, CE .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (06) :721-723
[6]   NEW GIANTURCO-GRIFKA VASCULAR OCCLUSION DEVICE - INITIAL STUDIES IN A CANINE MODEL [J].
GRIFKA, RG ;
MULLINS, CE ;
GIANTURCO, C ;
NIHILL, MR ;
OLAUGHLIN, MP ;
SLACK, MC ;
CLUBB, FJ ;
MYERS, TJ .
CIRCULATION, 1995, 91 (06) :1840-1846
[7]   Use of a self-expanding vascular occluder for embolization during endovascular aortic aneurysm repair [J].
Kato, N ;
Semba, CP ;
Dake, MD .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 8 (01) :27-33
[8]  
LIU X, 1990, J APPL MANUFACT SYS, V3, P65
[9]   A NEW DEVICE FOR TRANSCATHETER CLOSURE OF THE PATENT DUCTUS-ARTERIOSUS - A FEASIBILITY STUDY IN DOGS [J].
MAGAL, C ;
WRIGHT, KC ;
DUPRAT, G ;
WALLACE, S ;
GIANTURCO, C .
INVESTIGATIVE RADIOLOGY, 1989, 24 (04) :272-276
[10]   Catheter closure of moderate- to large-sized patent ductus arteriosus using the new Amplatzer duct occluder: Immediate and short-term results [J].
Masura, J ;
Walsh, KP ;
Thanopoulous, B ;
Chan, C ;
Bass, J ;
Goussous, Y ;
Gavora, P ;
Hijazi, ZM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (04) :878-882