Effects of stimulation mode, level and location on forward-masked excitation patterns in cochlear implant patients

被引:49
作者
Chatterjee, M [1 ]
Galvin, JJ [1 ]
Fu, QJ [1 ]
Shannon, RV [1 ]
机构
[1] House Ear Res Inst, Dept Auditory Implants & Percept, Los Angeles, CA 90057 USA
来源
JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY | 2006年 / 7卷 / 01期
关键词
D O I
10.1007/s10162-005-0019-2
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
In multi-channel cochlear implants, electrical current is delivered to appropriate electrodes in the cochlea to approximate the spatial representation of speech. Theoretically, electrode configurations that restrict the current spread within the cochlea (e.g., bi- or tripolar stimulation) may provide better spatial selectivity, and in turn, better speech recognition than configurations that produce a broader current spread (e.g., monopolar stimulation). However, the effects of electrode configuration on supra-threshold excitation patterns have not been systematically studied in cochlear implant patients. In the present study, forward-masked excitation patterns were measured in cochlear implant patients as functions of stimulation mode, level and location within the cochlea. All stimuli were 500 pulses-per-second biphasic pulse trains (200 mu s/phase, 20 mu s inter-phase gap). Masker stimuli were 200 ms in duration; the bi-polar configuration was varied from narrow (BP + 1) to wide (BP + 17), depending on the test condition. Probe stimuli were 20 ms in duration and the masker-probe delay was 5 ins; the probe configuration was fixed at BP + 1. The results indicated that as the distance between the active and return electrodes in a bi-polar pair was increased, the excitation pattern broadened within the cochlea. When the distance between active and return electrodes was sufficiently wide, two peaks were often observed in the excitation pattern, comparable to non-overlapping electric fields produced by widely separated dipoles. Analyses of the normalized data showed little effect of stimulation level on the shape of the excitation pattern.
引用
收藏
页码:15 / 25
页数:11
相关论文
共 24 条
[1]   Cortical responses to cochlear implant stimulation: Channel interactions [J].
Bierer, JA ;
Middlebrooks, JC .
JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2004, 5 (01) :32-48
[2]   Auditory cortical images of cochlear-implant stimuli: Dependence on electrode configuration [J].
Bierer, JA ;
Middlebrooks, JC .
JOURNAL OF NEUROPHYSIOLOGY, 2002, 87 (01) :478-492
[3]   Forward masking in different cochlear implant systems [J].
Boëx, C ;
Kós, MI ;
Pelizzone, M .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2003, 114 (04) :2058-2065
[4]   Electrically evoked whole nerve action potentials in ineraid cochlear implant users: Responses to different stimulating electrode configurations and comparison to psychophysical responses [J].
Brown, CJ ;
Abbas, PJ ;
Borland, J ;
Bertschy, MR .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1996, 39 (03) :453-467
[5]   Forward masked excitation patterns in multielectrode electrical stimulation [J].
Chatterjee, M ;
Shannon, RV .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 103 (05) :2565-2572
[6]   Temporal mechanisms underlying recovery from forward masking in multielectrode-implant listeners [J].
Chatterjee, M .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 105 (03) :1853-1863
[7]   Psychophysics of a prototype peri-modiolar cochlear implant electrode array [J].
Cohen, LT ;
Saunders, E ;
Clark, GM .
HEARING RESEARCH, 2001, 155 (1-2) :63-81
[8]   Effects of stimulus level on speech perception with cochlear prostheses [J].
Franck, KH ;
Xu, L ;
Pfingst, BE .
JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2003, 4 (01) :49-59
[9]   Speech recognition in noise as a function of the number of spectral channels: Comparison of acoustic hearing and cochlear implants [J].
Friesen, LM ;
Shannon, RV ;
Baskent, D ;
Wang, X .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2001, 110 (02) :1150-1163
[10]   Noise susceptibility of cochlear implant users: The role of spectral resolution and smearing [J].
Fu, QJ ;
Nogaki, G .
JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2005, 6 (01) :19-27