Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams

被引:63
作者
De Vlamynck, K
Palmans, H
Verhaegen, F
De Wagter, C
De Neve, W
Thierens, H
机构
[1] State Univ Ghent Hosp, Dept Radiotherapy, B-9000 Ghent, Belgium
[2] State Univ Ghent, Dept Biomed Phys, Stand Dosimetry Lab, B-9000 Ghent, Belgium
关键词
Monte Carlo; small fields; diamond detector; multileaf collimator (MLC);
D O I
10.1118/1.598693
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Small fields where electronic equilibrium is not achieved are becoming increasingly important in clinical practice. These complex situations give rise to problems and inaccuracies in both dosimetry and analytical/empirical dose calculation, and therefore require other than conventional methods. A natural diamond detector and a Markus parallel plate ionization chamber have been selected for clinical dosimetry in 6 MV photon beams. Results of simulations using the Monte Carlo system BEAM/EGS4 to model the beam geometry have been compared with dose measurements. A modification of the existing component module for multileaf collimators (MLCs) allowed the modeling of a linear accelerator SL 25 (Elekta Oncology Systems) equipped with a MLC with curved leaf-ends. A mechanical measurement method with spacer plates and a light-field edge detection technique are described as methods to obtain geometrical data of collimator openings for application in the Monte Carlo system. Generally a good agreement is found between measurements and calculations of depth dose distributions and deviations are typically less than 1%. Calculated lateral dose profiles slightly exceed measured dose distributions near the higher level of the penumbras for a 10 x 2 cm(2) held, but agree well with the measurements for all other cases. The simulations sue also able to predict variations of output factors and ratios of output factors as a function of field width and field-offset. The Monte Carlo results demonstrate that qualitative changes in energy spectra are toe, small to explain these variations and that especially geometrical factors affect the output factors and depth dose curves and profiles. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)01109-8].
引用
收藏
页码:1874 / 1882
页数:9
相关论文
共 26 条
[1]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[2]   CALCULATION AND APPLICATION OF POINT SPREAD FUNCTIONS FOR TREATMENT PLANNING WITH HIGH-ENERGY PHOTON BEAMS [J].
AHNESJO, A ;
ANDREO, P ;
BRAHME, A .
ACTA ONCOLOGICA, 1987, 26 (01) :49-56
[3]   DOSE MEASUREMENTS AND CALCULATIONS OF SMALL RADIATION-FIELDS FOR 9-MV X-RAYS [J].
ARCOVITO, G ;
PIERMATTEI, A ;
DABRAMO, G ;
BASSI, FA .
MEDICAL PHYSICS, 1985, 12 (06) :779-784
[4]  
AYYANGAR KM, 1996, MED PHYS, V23, P1129
[5]   ABSORBED DOSE PERTURBATION CAUSED BY DIODES FOR SMALL-FIELD PHOTON DOSIMETRY [J].
BEDDAR, AS ;
MASON, DJ ;
OBRIEN, PF .
MEDICAL PHYSICS, 1994, 21 (07) :1075-1079
[6]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[7]  
BJARNGARD BE, 1990, MED PHYS, V17, P794, DOI 10.1118/1.596475
[8]   CT-based Monte Carlo simulation tool for dosimetry planning and analysis [J].
DeMarco, JJ ;
Solberg, TD ;
Smathers, JB .
MEDICAL PHYSICS, 1998, 25 (01) :1-11
[9]   DOSIMETRY OF SMALL RADIATION-FIELDS FOR 10-MV X-RAYS [J].
HOUDEK, PV ;
VANBUREN, JM ;
FAYOS, JV .
MEDICAL PHYSICS, 1983, 10 (03) :333-336
[10]   THE DESIGN AND PERFORMANCE-CHARACTERISTICS OF A MULTILEAF COLLIMATOR [J].
JORDAN, TJ ;
WILLIAMS, PC .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (02) :231-251