Memory artifact related to selenium-based digital radiography systems

被引:18
作者
Chotas, HG [1 ]
Floyd, CE [1 ]
Ravin, CE [1 ]
机构
[1] DUKE UNIV, MED CTR, DEPT BIOMED ENGN, DURHAM, NC 27710 USA
关键词
radiography; digital; thorax;
D O I
10.1148/radiology.203.3.9169721
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Digital images acquired on radiography systems with amorphous selenium detectors are susceptible to ''memory artifacts'' from prior x-ray exposures. In routine clinical use and in a laboratory experiment, artifacts appeared in chest radiographs until the selenium recovered from initial exposure. Memory artifacts were eliminated when 3 minutes or more elapsed between acquisition of a lateral chest radiograph and acquisition of the next radiograph.