Effects of the sterilisation method on the wear of UHMWPE acetabular cups tested in a hip joint simulator

被引:54
作者
Affatato, S
Bordini, B
Fagnano, C
Taddei, P
Tinti, A
Toni, A
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Ctr Studio Spettroscopia Raman, Dipartimento Biochim G Moruzzi, Sezione Chim & Propedeut Biochim, I-40126 Bologna, Italy
[3] 1 Div Ortopedia & Traumatol, Bologna, Italy
关键词
ethylene oxide gas sterilisation; gamma irradiation; polyethylene sterilisation; in vitro wear; micro-Raman spectroscopy;
D O I
10.1016/S0142-9612(01)00265-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultra-high molecular-weight-polyethylene is the most commonly used bearing material in total joint replacement. Wear of polyethylene is a serious clinical problem that limits the longevity of orthopaedic implants. Information on degradative changes in the material properties and on the methods used for the sterilisation of polyethylene may help in the selection process of orthopaedic implants with the best wear resistance. This study was performed to investigate the effects of the sterilisation method (gamma irradiation and ethylene oxide treatment) on the wear and on the changes in physical properties of polyethylene acetabular cups. At this purpose, gamma-sterilised and ethylene oxide (EtO)-sterilised acetabular cups were tested against CoCr femoral heads in a hip joint simulator run for 5 million cycles in bovine calf serum. The crystallinity of the cups was evaluated by micro-Raman spectroscopy as a function of the inner surface position. The partial least square calibration was used to correlate the Raman spectra with the crystallinity of the polymer measured by differential scanning calorimetry. The analysis performed on soak control acetabular cups demonstrated that the gamma-sterilised cups are significantly more crystalline than the EtO-sterilised ones. The mean crystallinity values obtained for the gamma-sterilised and EtO-sterilised soak control cups were 65.0% and 63.4%, respectively. Weight loss measurements revealed that the gamma-sterilised acetabular cups exhibited a lower wear rate than that by EtO-sterilised. The Raman results obtained on gamma-sterilised and EtO-sterilised acetabular cups showed that the changes in surface crystallinity were mainly caused by irradiation rather than by the mechanical friction during the in vitro tests. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1439 / 1446
页数:8
相关论文
共 58 条
[51]  
SOMMERICH R, 1996, T ORTHOP RES SOC, V21, P486
[52]  
SUN DC, 1996, P 5 WORLD BIOM C, V1, P195
[53]  
WANG A, 1996, T 5 WORLD BIOM C, P198
[54]  
WEAVER KD, 1995, P 21 ANN M SOC BIOM, P129
[55]  
White SE, 1996, CLIN ORTHOP RELAT R, P164
[56]   USE OF MICRO-RAMAN SPECTROSCOPY FOR THE QUANTITATIVE-DETERMINATION OF POLYETHYLENE DENSITY USING PARTIAL LEAST-SQUARES CALIBRATION [J].
WILLIAMS, KPJ ;
EVERALL, NJ .
JOURNAL OF RAMAN SPECTROSCOPY, 1995, 26 (06) :427-433
[57]  
Witkiewicz H, 1996, J BIOMED MATER RES, V33, P73, DOI 10.1002/(SICI)1097-4636(199622)33:2<73::AID-JBM3>3.3.CO
[58]  
2-0