Determination of copper, cadmium and lead in biological samples by isotope dilution inductively coupled plasma mass spectrometry after on-line pre-treatment by anodic stripping voltammetry

被引:66
作者
Hwang, TJ [1 ]
Jiang, SJ [1 ]
机构
[1] NATL SUN YAT SEN UNIV,DEPT CHEM,KAOHSIUNG 804,TAIWAN
关键词
isotope dilution; inductively coupled plasma mass spectrometry; anodic stripping voltammetry; matrix separation; copper; cadmium; lend; biological sample;
D O I
10.1039/ja9961100353
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Flow injection isotope dilution ICP-MS was applied to the determination of Cu, Cd and Ph in several biological samples after on-line pre-treatment by anodic stripping voltammetry. The isotope ratios for each injection were calculated from the areas of the how injection peaks. The influence of the operating conditions of the electrochemical system on the analyte signals was studied. The influence of non-spectroscopic and spectroscopic interferences caused by the matrix on the precision and accuracy of isotope ratio determinations was also investigated, The precision for isotope ratio determinations was better than 2%, Detection limits were 0.07, 0.01 and 0.07 ng ml(-1) for Cn, Cd and Pb, respectively. The isotope dilution ICP-MS method was successfully applied to the determination of Co, Cd and Pb in National Research Council of Canada CRMs DORM-1 Dogfish Muscle and DOLT-1 Dogfish Liver and NIST SRM 2670 Toxic Metals in Freeze-Dried Urine.
引用
收藏
页码:353 / 357
页数:5
相关论文
共 19 条
[1]   MEASUREMENT OF VANADIUM, NICKEL, AND ARSENIC IN SEAWATER AND URINE REFERENCE MATERIALS BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY WITH CRYOGENIC DESOLVATION [J].
ALVES, LC ;
ALLEN, LA ;
HOUK, RS .
ANALYTICAL CHEMISTRY, 1993, 65 (18) :2468-2471
[2]   ADVANCED ELECTROANALYTICAL TECHNIQUES VERSUS ATOMIC-ABSORPTION SPECTROMETRY, INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY AND INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY IN ENVIRONMENTAL-ANALYSIS [J].
BERSIER, PM ;
HOWELL, J ;
BRUNTLETT, C .
ANALYST, 1994, 119 (02) :219-232
[3]   EVALUATION OF ETHENE ADDITION TO THE NEBULIZER GAS IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY FOR THE REMOVAL OF MATRIX-GAS-DERIVED, SOLVENT-GAS-DERIVED, AND SUPPORT-GAS-DERIVED POLYATOMIC ION INTERFERENCES [J].
EBDON, L ;
FORD, MJ ;
HUTTON, RC ;
HILL, SJ .
APPLIED SPECTROSCOPY, 1994, 48 (04) :507-516
[4]   SIMPLE APPROACH TO REDUCING POLYATOMIC ION INTERFERENCES ON ARSENIC AND SELENIUM IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EVANS, EH ;
EBDON, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (03) :299-300
[5]   MASS-SPECTROMETRY OF INDUCTIVELY COUPLED PLASMAS [J].
HOUK, RS .
ANALYTICAL CHEMISTRY, 1986, 58 (01) :A97-&
[6]   DETERMINATION OF TRACE LEVELS OF METAL-IONS IN WATER SAMPLES BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY AFTER ONLINE PRECONCENTRATION ON SO3-OXINE CM-CELLULOSE [J].
HUANG, KS ;
JIANG, SJ .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 347 (6-7) :238-242
[7]  
Jarvis K. E., 1992, HDB INDUCTIVELY COUP, DOI [DOI 10.1007/BF02290251, 10.1007/BF02290251]
[8]  
KISSINGER PT, 1984, LABORATORY TECHNIQUE
[9]   REDUCTION OF POLYATOMIC INTERFERENCES IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY BY SELECTION OF INSTRUMENTAL PARAMETERS AND USING AN ARGON-NITROGEN PLASMA - EFFECT ON MULTIELEMENT ANALYSES [J].
LABORDA, F ;
BAXTER, MJ ;
CREWS, HM ;
DENNIS, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (06) :727-736
[10]   DETERMINATION OF TRACES OF COPPER, CADMIUM AND LEAD IN BIOLOGICAL AND ENVIRONMENTAL-SAMPLES BY FLOW-INJECTION ISOTOPE-DILUTION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
LU, PL ;
HUANG, KS ;
JIANG, SJ .
ANALYTICA CHIMICA ACTA, 1993, 284 (01) :181-188