Optimization of the atmospheric pressure chemical ionization liquid chromatography mass spectrometry interface

被引:45
作者
Garcia, DM [1 ]
Huang, SK [1 ]
Stansbury, WF [1 ]
机构
[1] HOECHST CELANESE CORP,CORPUS CHRISTI TECH CTR,DEPT ANALYT,CORPUS CHRISTI,TX 78469
关键词
D O I
10.1016/1044-0305(95)00620-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Use of optimized instrument parameters that result from statistical experimentation revealed that the sensitivity of atmospheric pressure chemical ionization (APCI) liquid chromatography-mass spectrometry (LC/MS) is greater than the sensitivity of an optimized Thermabeam(TM) LC/MS interface by about 3 orders of magnitude, when tested on aromatic compounds. APCI is one of the few LC/MS techniques in which the chromatogram is directly comparable with liquid chromatographs that use ultraviolet detection. The optimum flow rates (e.g., 0.1 mL/min), relatively low capillary heat (e.g., 225 degrees C), and high sheath-gas pressure (e.g., 60 lb/in(2)). The optimization was achieved by monitoring the responses of sensitivity, fragmentation, and cluster ion formation. The fine tuning for high sensitivity calls for a high percentage of water in the mobile phase. In contrast, a high percentage of organic content in the mobile phase is required to obtain abundant protonated molecular ions with respect to fragmentation and clustering. This is an important consideration for analyses of unknowns.
引用
收藏
页码:59 / 65
页数:7
相关论文
共 29 条
[11]   OPTIMIZATION OF A COMMERCIAL LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY INTERFACE BY STATISTICAL EXPERIMENTAL-DESIGN [J].
HUANG, SK ;
GLANCY, SB ;
STANSBURY, WF .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1993, 7 (08) :722-724
[12]  
HUERTAS ML, 1975, ATMOS ENVIRON, V9, P1018, DOI 10.1016/0004-6981(75)90023-2
[13]   ELECTROSPRAY ION SPRAY - A COMPARISON OF MECHANISMS AND PERFORMANCE [J].
IKONOMOU, MG ;
BLADES, AT ;
KEBARLE, P .
ANALYTICAL CHEMISTRY, 1991, 63 (18) :1989-1998
[14]   FROM IONS IN SOLUTION TO IONS IN THE GAS-PHASE - THE MECHANISM OF ELECTROSPRAY MASS-SPECTROMETRY [J].
KEBARLE, P ;
TANG, L .
ANALYTICAL CHEMISTRY, 1993, 65 (22) :A972-A986
[15]  
MARK H, 1994, SPECTROSCOPY, V9, P30
[16]  
Mark H., 1994, SPECTROSCOPY, V9, P26
[17]  
MARK H, 1994, SPECTROSCOPY, V10, P17
[18]  
Montgomery D. C., 1991, DESIGN ANAL EXPT
[19]   CHEMICAL DERIVATIZATION FOR ELECTROSPRAY-IONIZATION MASS-SPECTROMETRY .1. ALKYL-HALIDES, ALCOHOLS, PHENOLS, THIOLS, AND AMINES [J].
QUIRKE, JME ;
ADAMS, CL ;
VANBERKEL, GJ .
ANALYTICAL CHEMISTRY, 1994, 66 (08) :1302-1315
[20]   NEGATIVE-ION FORMATION IN ELECTROSPRAY MASS-SPECTROMETRY [J].
STRAUB, RF ;
VOYKSNER, RD .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1993, 4 (07) :578-587