Automated feedback control of body temperature for small animal studies with MR microscopy

被引:21
作者
Qiu, HH [1 ]
Cofer, GP [1 ]
Hedlund, LW [1 ]
Johnson, GA [1 ]
机构
[1] DUKE UNIV,MED CTR,DEPT RADIOL,CTR INVIVO MICROSCOPY,DURHAM,NC 27710
关键词
body temperature control; computer control; feedback control; magnetic resonance microscopy; rodents;
D O I
10.1109/10.641338
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A temperature control system consisting of a thermistor, signal processor, and computer algorithm was developed for magnetic resonance (MR) microscopy of small live animals. With control of body temperature within +/-0.2 degrees C of the set point, heart rate is stabilized and, in turn, repetition time (TR) during cardiac-gated studies is less variable. Thus, image quality and resolution are improved.
引用
收藏
页码:1107 / 1113
页数:7
相关论文
共 17 条
[1]  
[Anonymous], APPL MAGNETIC RESONA
[2]  
[Anonymous], REV MAGN RESON MED
[3]   DIFFUSION-WEIGHTED MR MICROSCOPY WITH FAST SPIN-ECHO [J].
BEAULIEU, CF ;
ZHOU, XH ;
COFER, GP ;
JOHNSON, GA .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (02) :201-206
[4]   In vivo He-3 MR images of guinea pig lungs [J].
Black, RD ;
Middleton, HL ;
Cates, GD ;
Cofer, GP ;
Driehuys, B ;
Happer, W ;
Hedlund, LW ;
Johnson, GA ;
Shattuck, MD ;
Swartz, JC .
RADIOLOGY, 1996, 199 (03) :867-870
[5]   A new in vivo method for quantitative analysis of stroke lesions using diffusion-weighted magnetic resonance microscopy [J].
Hall, WL ;
Benveniste, H ;
Hedlund, LW ;
Johnson, GA .
NEUROIMAGE, 1996, 3 (03) :158-166
[6]  
HEDLUND L, 1995, P MR MICR C WURZB GE, P36
[7]   A VENTILATOR FOR MAGNETIC-RESONANCE-IMAGING [J].
HEDLUND, LW ;
DEITZ, J ;
NASSAR, R ;
HERFKENS, R ;
VOCK, P ;
DAHLKE, J ;
KUBEK, R ;
EFFMANN, EL ;
PUTMAN, CE .
INVESTIGATIVE RADIOLOGY, 1986, 21 (01) :18-23
[8]  
HEDLUND LW, 1991, P ASS U RAD ANN M OR
[9]  
JOHNSON GA, 1993, MAGN RESON QUART, V9, P1
[10]   THEORETICAL AND EXPERIMENTAL EVALUATION OF A TEMPERATURE CONTROLLER FOR SCANNED FOCUSED ULTRASOUND HYPERTHERMIA [J].
LIN, WL ;
ROEMER, RB ;
HYNYNEN, K .
MEDICAL PHYSICS, 1990, 17 (04) :615-625