共 4 条
Interferometric fiber-based optical biosensor to measure ultra-small changes in refractive index
被引:11
作者:
Akkin, T
[1
]
Davé, DP
[1
]
Milner, TE
[1
]
Rylander, HG
[1
]
机构:
[1] Univ Texas, Austin, TX 78712 USA
来源:
OPTICAL FIBERS AND SENSORS FOR MEDICAL APPLICATIONS II
|
2002年
/
4616卷
关键词:
refractive index;
birefringence;
biosensor;
phase sensitive;
fiber;
interferometer;
D O I:
10.1117/12.463805
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
In many bio-medical optical imaging applications such as monitoring cell and tissue dynamics and imaging of phase objects, ultra-small changes in refractive index must be detected. We describe a fiber-based optical biosensor, which is capable of detecting ultra-small refractive index changes in highly scattering media with high lateral and longitudinal spatial resolution. The system is a dual channel phase-sensitive optical low coherence tomography system that measures relative optical path length differences between the orthogonal modes of the polarization-maintaining fiber.