A poly(vinyl chloride) membrane electrode for the determination of the diuretic furosemide

被引:27
作者
Dias, ILT
Neto, GD
Vendramini, DC
Sommer, C
Martins, JLS
Kubota, LT
机构
[1] Univ Sao Francisco, UAACBS, Curso Farm, BR-12916900 Braganca Paulista, SP, Brazil
[2] Univ Sao Paulo, Fac Ciencias Farmaceut, Sao Paulo, Brazil
[3] Univ Estadual Campinas, Inst Quim, Campinas, SP, Brazil
关键词
furosemide; poly(vinyl chloride) membrane electrode; ion-selective electrode; direct potentiometry; pharmaceutical analysis;
D O I
10.1081/AL-120027772
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
New polymeric electrode responding to furosemide was constructed by incorporating the furosemide-aliquat ion pair complex into a poly(vinyl chloride) (PVC) membrane. The electrode exhibited a linear response over the concentration range 1.59 x 10(-4)-1.00 X 10(-2) mol L-1, a detection limit Of,. 19 X 10(-4) mol L-1 with E = - 245.6 - 58.9log[Fur]. The pH did not affect the response of the electrode within the range 7.0-10.0. Direct potentiometry was used to assay furosemide in pharmaceutical preparations. The results were compared with those obtained with spectrophotometric and chromatographic reference methods.
引用
收藏
页码:35 / 46
页数:12
相关论文
共 44 条
[21]  
IUPAC, 1979, PURE APPL CHEM, V48, P127, DOI DOI 10.1351/PAC197648010127
[22]   Piroxicam and tenoxicam selective membrane sensors [J].
Khalil, S ;
Borham, N ;
El-Ries, MA .
ANALYTICA CHIMICA ACTA, 2000, 414 (1-2) :215-219
[23]  
LIGTH T, 1997, J CHEM EDUC, V74, P171
[24]  
MAKMOUD WA, 2001, ANAL CHIM ACTA, V436, P199
[25]   Poly(vinyl chloride) membrane electrodes for the determination of the antibiotic tiamulin [J].
Marques, MRC ;
Neto, GDO .
ANALYTICAL SCIENCES, 1997, 13 (03) :457-461
[26]  
NIKOLIC K, 1990, ACTA PHARM JUGOSL, V40, P521
[27]  
NIKOLIC KI, 1989, J PHARM BELG, V44, P387
[28]  
PANCHAGNULA R, 1997, PHARM SCI, V3, P425, DOI DOI 10.1111/J.2042-7158.1997.TB00296.X
[29]  
PONTO LLB, 1990, CLIN PHARMACOKINET, V18, P381
[30]  
PONTO LLB, 1990, CLIN PHARMACOKINET, V18, P460