Automatic cardiac LV boundary detection and tracking using hybrid fuzzy temporal and fuzzy multiscale edge detection

被引:36
作者
Setarehdan, SK [1 ]
Soraghan, JJ [1 ]
机构
[1] Univ Strathclyde, Signal Proc Div, Glasgow G1 1XW, Lanark, Scotland
关键词
boundary detection; echocardiography; edge detection; fuzzy logic; wavelet transform;
D O I
10.1109/10.797997
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes a new fully automatic fuzzy multiresolution-based algorithm for cardiac left ventricular (LV) epicardial and endocardial boundary detection and tracking on a sequence of short axis (SA) echocardiographic images of a complete cardiac cycle. This is a necessary step for automatic quantification of cardiac function using echo images. The proposed method is a "center-based" approach in which epicardial and endocardial boundary edge points are searched for on radial lines emanating from the LV center point. The central point of the LV cavity is estimated using a fuzzy-based technique in which the "uncertain" spatial, morphological, and intensity information of the image are represented as fuzzy sets and then combined by fuzzy operators. Edge-detection stage uses multiscale spatial and temporal information in a fuzzy multiresolution framework to identify a single moving edge point for each one of the epicardial and endocardial boundaries over the M radii in the N frames of a complete cardiac cycle. The ram extracted edge points are then processed in the wavelet domain to reduce the effects of noise from the boundaries and papillary muscles from the endocardial boundary extraction process. Finally, a uniform cubic B-spline approximation method is used to define the closed LV boundaries. Experiments with simulated and red echocardiographic images are presented.
引用
收藏
页码:1364 / 1378
页数:15
相关论文
共 40 条
[11]   RECOVERY OF THE 3-D SHAPE OF THE LEFT-VENTRICLE FROM ECHOCARDIOGRAPHIC IMAGES [J].
COPPINI, G ;
POLI, R ;
VALLI, G .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (02) :301-317
[12]   MATCHED-FILTER IDENTIFICATION OF LEFT-VENTRICULAR ENDOCARDIAL BORDERS IN TRANSESOPHAGEAL ECHOCARDIOGRAMS [J].
DETMER, PR ;
BASHEIN, G ;
MARTIN, RW .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (04) :396-404
[13]   Wall position and thickness estimation from sequences of echocardiographic images [J].
Dias, JMB ;
Leitao, JMN .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (01) :25-38
[14]  
Dubois D.J., 1980, FUZZY SETS SYSTEMS T
[15]   EPICARDIAL BOUNDARY DETECTION USING FUZZY-REASONING [J].
FENG, J ;
LIN, WC ;
CHEN, CT .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (02) :187-199
[16]   AUTOMATIC VENTRICULAR CAVITY BOUNDARY DETECTION FROM SEQUENTIAL ULTRASOUND IMAGES USING SIMULATED ANNEALING [J].
FRIEDLAND, N ;
ADAM, D .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (04) :344-353
[17]   KNOWLEDGE-BASED IMAGE-ANALYSIS FOR AUTOMATED BOUNDARY EXTRACTION OF TRANSESOPHAGEAL ECHOCARDIOGRAPHIC LEFT-VENTRICULAR IMAGES [J].
HAN, CY ;
LIN, KN ;
WEE, WG ;
MINTZ, RM ;
POREMBKA, DT .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (04) :602-610
[18]  
Hunter IA, 1995, COMPUT CARDIOL, P741, DOI 10.1109/CIC.1995.482771
[19]   Automatic identification of papillary muscles in left-ventricular short-axis echocardiographic images [J].
Karras, T ;
Wilson, DC ;
Geiser, EA ;
Conetta, DA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (05) :460-470
[20]  
KELLER JM, 1992, INTRO FUZZY LOGIC AP, P122