Limited range of motion of hip resurfacing arthroplasty due to unfavorable ratio of prosthetic head size and femoral neck diameter

被引:49
作者
Kluess, Daniel [1 ,2 ]
Zietz, Carmen [1 ]
Lindner, Tobias [1 ]
Mittelmeier, Wolfram [1 ]
Schmitz, Klaus-Peter [2 ]
Bader, Rainer [1 ]
机构
[1] Univ Rostock, Dept Orthoped, D-2500 Rostock 1, Germany
[2] Univ Rostock, Inst Biomed Engn, D-2500 Rostock 1, Germany
关键词
D O I
10.1080/17453670810016803
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and purpose Hip resurfacing arthroplasty is being used more and more frequently. The small ratio in size between the resurfaced femoral head and the relatively thick femoral neck raises the question of whether the range of motion is sufficient, particularly with regard to the high mobility required by younger patients. We analyzed motion in a CAD model. Methods Three-dimensional CAD models of the natural hip were created from CT scans and 8 designs of hip resurfacing prostheses (head diameter between 42 mm and 54 mm combined with a hemispherical cup) were implanted in a virtual sense. We simulated 3 different leg positions and the range of motion was evaluated, considering five different implant positions. Results The range of motion of the hip resurfacing designs analyzed was far below the range of motion of stemmed total hip prostheses. None of the resurfacing prostheses provided flexion movements of 90 degrees without impingement. The average range of motion of hip resurfacing arthroplasty was 31-48 below the range of motion of a stemmed total hip replacement with 32-mm head diameter. Interpretation The range of motion of the hip resurfacing designs examined was substantially less than that of a conventional total hip prosthesis. Since impingement of the femoral neck on the acetabular component increases the risk of neck fractures, of dislocation and of subsequent implant loosening, the design and position of the implant should be considered before using hip resurfacing arthroplasty as a standard treatment for younger patients.
引用
收藏
页码:748 / 754
页数:7
相关论文
共 28 条
[1]
Amstutz Harlan C, 2006, J Bone Joint Surg Am, V88 Suppl 1 Pt 2, P234, DOI 10.2106/JBJS.F.00273
[2]
Results of metal-on-metal hybrid hip resurfacing for Crowe type-I and II developmental dysplasia [J].
Amstutz, Harlan C. ;
Antoniades, John T. ;
Le Duff, Michel J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (02) :339-346
[3]
Metal-on-metal hybrid surface arthroplasty:: Two to six-year follow-up study [J].
Amstutz, HC ;
Beaulé, PE ;
Dorey, FJ ;
Le Duff, MJ ;
Campbell, PA ;
Gruen, TA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01) :28-39
[4]
Fracture of the neck of the femur after surface arthroplasty of the hip [J].
Amstutz, HC ;
Campbell, PA ;
Le Duff, MJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (09) :1874-1877
[5]
The influence of head and neck geometry on stability of total hip replacement - A mechanical test study [J].
Bader, R ;
Scholz, R ;
Steinhauser, E ;
Zimmermann, S ;
Busch, R ;
Mittelmeier, W .
ACTA ORTHOPAEDICA SCANDINAVICA, 2004, 75 (04) :415-421
[6]
Method for the evaluation of factors influencing the dislocation stability of total hip endoprotheses [J].
Bader, R ;
Scholz, R ;
Steinhauser, E ;
Busch, R ;
Mittlemeier, W .
BIOMEDIZINISCHE TECHNIK, 2004, 49 (05) :137-144
[7]
Computer-based motion simulation of total hip prostheses with ceramic-on-ceramic wear couple. Analysis of implant design and orientation as influence parameters [J].
Bader, R ;
Steinhauser, E ;
Gradinger, R ;
Willmann, G ;
Mittelmeier, W .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2002, 140 (03) :310-316
[8]
Vascularity of the arthritic femoral head and hip resurfacing [J].
Beaule, Paul E. ;
Campbell, Pat ;
Lu, Zhen ;
Leunig-Ganz, Katharina ;
Beck, Martin ;
Leunig, Michael ;
Ganz, Reinhold .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A :85-96
[9]
Biomechanics of large femoral heads - What they do and don't do [J].
Crowninshield, RD ;
Maloney, WJ ;
Wentz, DH ;
Humphrey, SM ;
Blanchard, CR .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :102-107
[10]
Resurfacing hip prostheses revisited - Failure analysis during a 16-year follow-up [J].
Duijsens, AWHB ;
Keizer, S ;
Vliet-Vlieland, T ;
Nelissen, RGHH .
INTERNATIONAL ORTHOPAEDICS, 2005, 29 (04) :224-228