Imaging columns of the light elements carbon, nitrogen and oxygen with sub Angstrom resolution

被引:108
作者
Kisielowski, C
Hetherington, CJD
Wang, YC
Kilaas, R
O'Keefe, MA
Thust, A
机构
[1] Univ Calif Berkeley, Lawrence Berkeley Lab, Natl Ctr Electron Microscopy, Div Sci Mat, Berkeley, CA 94720 USA
[2] Univ Oxford, Dept Mat, Oxford OX1 3PH, England
[3] FEI Co, Hillsboro, OR 97124 USA
[4] Forschungszentrum Julich, Inst Festkorperforsch, D-5242 Julich, Germany
基金
美国国家科学基金会;
关键词
D O I
10.1016/S0304-3991(01)00090-0
中图分类号
TH742 [显微镜];
学科分类号
摘要
It is reported that lattice imaging with a 300 kV field emission microscope in combination with numerical reconstruction procedures can be used to reach an interpretable resolution of about 80 pm for the first time. A retrieval of the electron exit wave from focal series allows for the resolution of single atomic columns of the light elements carbon, nitrogen, and oxygen at a projected nearest neighbor spacing down to 85 pm. Lens aberrations are corrected on-line during the experiment and by hardware such that resulting image distortions are below 80 pm. Consequently, the imaging can be aberration-free to this extent. The resolution enhancement results from increased electrical and mechanical stability of the instrument coupled with a low spherical aberration coefficient of 0.595 + 0.005 mm. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:243 / 263
页数:21
相关论文
共 51 条
[1]  
[Anonymous], TRANSMISSION ELECT M
[2]  
BRANA A, 1998, ULTRAMICROSCOPY, V70, P16
[3]   PHASE RETRIEVAL THROUGH FOCUS VARIATION FOR ULTRA-RESOLUTION IN FIELD-EMISSION TRANSMISSION ELECTRON-MICROSCOPY [J].
COENE, W ;
JANSSEN, G ;
DEBEECK, MO ;
VANDYCK, D .
PHYSICAL REVIEW LETTERS, 1992, 69 (26) :3743-3746
[4]   Maximum-likelihood method for focus-variation image reconstruction in high resolution transmission electron microscopy [J].
Coene, WMJ ;
Thust, A ;
deBeeck, M ;
VanDyck, D .
ULTRAMICROSCOPY, 1996, 64 (1-4) :109-135
[5]  
COENE WMJ, 1992, 50 ANN P EMSA BOST M, P986
[6]   Wave function reconstruction in HRTEM: The parabola method [J].
deBeeck, MO ;
VanDyck, D ;
Coene, W .
ULTRAMICROSCOPY, 1996, 64 (1-4) :167-183
[7]  
*ERN ORL LAWR BERK, 1999, PUB826 ERN ORL LAWR
[8]   A spherical-aberration-corrected 200 kV transmission electron microscope [J].
Haider, M ;
Rose, H ;
Uhlemann, S ;
Schwan, E ;
Kabius, B ;
Urban, K .
ULTRAMICROSCOPY, 1998, 75 (01) :53-60
[9]   Image deconvolution for defected crystals in field-emission high-resolution electron microscopy [J].
He, WZ ;
Li, FH ;
Chen, H ;
Kawasaki, K ;
Oikawa, T .
ULTRAMICROSCOPY, 1997, 70 (1-2) :1-11
[10]  
ICHINOSE H, 1994, UNPUB INT WORKSH HIG