Microbore high-performance liquid chromatographic determination of cisapride in rat serum samples using column switching

被引:19
作者
Lee, HM [1 ]
Choi, SJ
Jeong, CK
Kim, YS
Lee, KC
Lee, HS
机构
[1] Wonkwang Univ, Coll Pharm, Iksan 570749, South Korea
[2] Sungkyunkwan Univ, Coll Pharm, Drug Targeting Lab, Suwon 440746, South Korea
来源
JOURNAL OF CHROMATOGRAPHY B | 1999年 / 727卷 / 1-2期
关键词
cisapride;
D O I
10.1016/S0378-4347(99)00103-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
For the determination of cisapride from serum samples, an automated microbore high-performance liquid chromatographic method with column switching has been developed. After serum samples (100 mu l) were directly injected onto a Capcell Pak MF Ph-1 pre-column (10x4 mm I.D.), the deproteinization and concentration were carried out by acetonitrile-phosphate buffer (20 mM, pH 7.0) (2:8, v/v) at valve position A. At 2.6 min, the valve was switched to position B and the concentrated analytes were transferred from MF ph-1 pre-column to a C-18 intermediate column (35x2 mm I.D.) using washing solvent. By valve switching to position A at 4.3 min, the analytes were separated on a Capcell Pak C-18 UG 120 column (250x1.5 mm I.D.) with acetonitrile-phosphate buffer (20 mM, pH 7.0) (5:5, v/v) at a how-rate of 0.1 ml/min. Total analysis time per sample was 18 min. The linearity of response was good (r=0.999) over the concentration range of 5-200 ng/ml. The within-day and day-to-day precision (CV) and inaccuracy were less than 3.7% and 3.8%, respectively. The mean recovery was 96.5+/-2.4% with the detection limit of 2 ng/ml. (C) 1999 Elsevier Science BN. All rights reserved.
引用
收藏
页码:213 / 217
页数:5
相关论文
共 13 条
[1]   Validation of chromatographic methods in biomedical analysis - Viewpoint and discussion [J].
Causon, R .
JOURNAL OF CHROMATOGRAPHY B, 1997, 689 (01) :175-180
[2]   Determination of cisapride and norcisapride in human plasma using high-performance liquid chromatography with ultraviolet detection [J].
Cisternino, S ;
Schlatter, J ;
Saulnier, JL .
JOURNAL OF CHROMATOGRAPHY B, 1998, 714 (02) :395-398
[3]  
GELDOF H, 1993, ALIMENT PHARM THERAP, V7, P409
[4]  
HUTTEMANN A, 1986, DIGESTION, V34, P143
[5]  
JANISCH HD, 1988, HEPATO-GASTROENTEROL, V35, P125
[6]   SYNTHESIS OF POLYMER-COATED MIXED-FUNCTIONAL PACKING MATERIALS FOR DIRECT ANALYSIS OF DRUG-CONTAINING SERUM AND PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
KANDA, T ;
KUTSUNA, H ;
OHTSU, Y ;
YAMAGUCHI, M .
JOURNAL OF CHROMATOGRAPHY A, 1994, 672 (1-2) :51-57
[7]   Simultaneous determination of losartan and active metabolite EXP3174 in rat plasma by HPLC with column switching [J].
Lee, H ;
Shim, HO ;
Lee, HS .
CHROMATOGRAPHIA, 1996, 42 (1-2) :39-42
[8]   On-line sample preparation of paraquat in human serum samples using high-performance liquid chromatography with column switching [J].
Lee, HS ;
Kim, K ;
Kim, JH ;
Do, KS ;
Lee, SK .
JOURNAL OF CHROMATOGRAPHY B, 1998, 716 (1-2) :371-374
[9]   Determination of myristicin in rat serum samples using microbore high performance liquid chromatography with column-switching [J].
Lee, HS ;
Kim, JH ;
Kim, K ;
Do, KS .
CHROMATOGRAPHIA, 1998, 48 (5-6) :365-368
[10]   SIMULTANEOUS DETERMINATION OF AMPICILLIN AND METAMPICILLIN IN BIOLOGICAL-FLUIDS USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH COLUMN-SWITCHING [J].
LEE, HY ;
LEE, JS ;
LEE, HS .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1995, 664 (02) :335-340