Evaluation of gradient inhomogeneity in the optimal design of gradient coils

被引:5
作者
Du, YP [1 ]
Parker, DL [1 ]
机构
[1] UNIV UTAH,HLTH SCI CTR,DEPT RADIOL,SALT LAKE CITY,UT 84132
关键词
gradient inhomogeneity; optimal design; gradient coils;
D O I
10.1016/0730-725X(95)02055-X
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The image distortion caused by gradient inhomogeneity was derived from the image equation of Fourier MR imaging, A measurement of gradient inhomogeneity was proposed based on the amount of image distortion. Using the proposed measurement of gradient inhomogeneity in the optimal design of a z-gradient coil reduced image distortion by 39-45% compared to a previously used figure of merit in gradient coil design.
引用
收藏
页码:201 / 207
页数:7
相关论文
共 11 条
[1]   THE LIMITATIONS OF A TARGET FIELD APPROACH TO COIL DESIGN [J].
ENGELSBERG, M ;
DESOUZA, RE ;
PAZOS, CMD .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1988, 21 (07) :1062-1066
[2]   OPTIMIZATION BY SIMULATED ANNEALING [J].
KIRKPATRICK, S ;
GELATT, CD ;
VECCHI, MP .
SCIENCE, 1983, 220 (4598) :671-680
[3]   NMR IMAGING IN THE PRESENCE OF MAGNETIC-FIELD INHOMOGENEITIES AND GRADIENT FIELD NONLINEARITIES [J].
ODONNELL, M ;
EDELSTEIN, WA .
MEDICAL PHYSICS, 1985, 12 (01) :20-26
[4]   SIGNAL-TO-NOISE EFFICIENCY IN MAGNETIC-RESONANCE-IMAGING [J].
PARKER, DL ;
GULLBERG, GT .
MEDICAL PHYSICS, 1990, 17 (02) :250-257
[5]   MINIMUM ENERGY MRI GRADIENT COILS OF GENERAL GEOMETRY [J].
PISSANETZKY, S .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1992, 3 (07) :667-673
[6]   IMAGE-RESTORATION FROM NON-UNIFORM MAGNETIC-FIELD INFLUENCE FOR DIRECT FOURIER NMR IMAGING [J].
SEKIHARA, K ;
KURODA, M ;
KOHNO, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (01) :15-24
[7]   MINIMUM INDUCTANCE COILS [J].
TURNER, R .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1988, 21 (10) :948-952
[8]   GRADIENT COIL DESIGN - A REVIEW OF METHODS [J].
TURNER, R .
MAGNETIC RESONANCE IMAGING, 1993, 11 (07) :903-920
[9]   SIMULATION OF THE INFLUENCE OF MAGNETIC-FIELD INHOMOGENEITY AND DISTORTION CORRECTION IN MR IMAGING [J].
WEIS, J ;
BUDINSKY, L .
MAGNETIC RESONANCE IMAGING, 1990, 8 (04) :483-489
[10]   COIL OPTIMIZATION FOR MRI BY CONJUGATE-GRADIENT DESCENT [J].
WONG, EC ;
JESMANOWICZ, A ;
HYDE, JS .