Blood flow in a continuous flow ventricular assist device

被引:33
作者
Allaire, PE
Wood, HG
Awad, RS
Olsen, DB
机构
[1] Univ Virginia, Dept Mech Aerosp & Nucl Engn, Charlottesville, VA USA
[2] Univ Virginia, Dept Biomed Engn, Charlottesville, VA USA
[3] Univ Utah, Artificial Heart Lab, Salt Lake City, UT USA
关键词
centrifugal blood pump; computational fluid dynamics; assisted circulation;
D O I
10.1046/j.1525-1594.1999.06420.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A numerical analysis was performed to predict the shear stresses, flow rates, and the velocity profiles in a continuous flow ventricular assist device, the CFVAD3. The problem was modeled as a rotating disk over a stationary disk. A variety of clearances was tested for the CFVAD3 coupled with a range of rotational speeds and pressure gradients. Velocity fields were generated using solutions obtained with FLOW3D software (AEA Technology, Pittsburgh, PA, U.S.A.) Analysis of these solutions shows that the pressure differential effect has a stronger influence on the flow than the rotational effect of the impeller Ekman layer. The predicted shear stresses reflect these changes in the volume flow rates and the speeds shown in the velocity profiles. Based on the predictions of the software, the optimum clearance and rotational speed were chosen. The conclusion is that a speed in the range of 2,200-2,400 rpm should be chosen depending on the efficiency of the pump.
引用
收藏
页码:769 / 773
页数:5
相关论文
共 24 条
[1]   Prototype continuous flow ventricular assist device supported on magnetic bearings [J].
Allaire, PE ;
Kim, HC ;
Maslen, EH ;
Olsen, DB ;
Bearnson, GB .
ARTIFICIAL ORGANS, 1996, 20 (06) :582-590
[2]   COMPUTATIONAL FLOW OPTIMIZATION OF ROTARY BLOOD PUMP COMPONENTS [J].
ANTAKI, JF ;
GHATTAS, O ;
BURGREEN, GW ;
HE, BC .
ARTIFICIAL ORGANS, 1995, 19 (07) :608-615
[3]   Pulsatile operation of a centrifugal ventricular assist device with magnetic bearings [J].
Bearnson, GB ;
Olsen, DB ;
Khanwilkar, PS ;
Long, JW ;
Allaire, PE ;
Maslen, EH .
ASAIO JOURNAL, 1996, 42 (05) :M620-M624
[4]   Development of a prototype magnetically suspended rotor ventricular assist device [J].
Bearnson, GB ;
Maslen, EH ;
Olsen, DB ;
Allaire, PE ;
Khanwilkar, PS ;
Long, JW ;
Kim, HC .
ASAIO JOURNAL, 1996, 42 (04) :275-281
[5]  
BLACKSHEAR PL, 1966, T AM SOC ART INT ORG, V12, P113
[6]  
BLACKSHEAR PL, 1987, HDB BIOENGINEERING, P501
[7]   MODEL FOR A GENERAL MECHANICAL BLOOD DAMAGE PREDICTION [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :583-589
[8]  
COONEY DO, 1976, BIOMEDICAL ENG PRINC, P51
[9]  
FUNG YC, 1984, BIODYNAMICS CIRCULAT, P64
[10]  
KANWILKAR PS, 1996, ARTIF ORGANS, V20, P597