Comparison of contrast-to-noise ratios for Bayesian processing and grids in digital chest radiography

被引:2
作者
Baydush, AH [1 ]
Gehm, WC [1 ]
Floyd, CE [1 ]
机构
[1] Duke Univ, Med Ctr, Durham, NC 27710 USA
来源
MEDICAL IMAGING 1999: IMAGE PROCESSING, PTS 1 AND 2 | 1999年 / 3661卷
关键词
Bayesian statistics; digital chest radiography; image processing; scatter;
D O I
10.1117/12.348622
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Previously, we have demonstrated the ability of Bayesian image estimation (BIE) to reduce scatter and improve image contrast to noise ratio (CNR) in chest radiography without degradation of resolution. Here, we compare the effectiveness of BIE to a standard 12:1 grid. Images of a geometric phantom with two inches of added polystyrene were obtained both with and without a 12:1, 150 Ip/mm grid. Images were acquired with standard protocols: 120 kVp, 72 inch source to image distance, and PA positioning. Images were acquired on a calibrated photostimulable phosphor system. An image exposure was used corresponding to the same patient dose as when acquiring film/screen chest images using a phototimer. The image acquired without the grid was processed by BIE for 6 iterations. Contrast, noise, and CNR were calculated and compared for the image acquired with the grid and the BIE processed image in different regions. BIE processing improved image CNR by 200 to 350% over that provided by the anti-scatter grid for the different regions. BIE provides higher CNR than that of a 12:1 grid. Because of this increase in CNR, Bayesian processed images will show an increase in detectability of low contrast objects, such as subtle lung nodules.
引用
收藏
页码:665 / 672
页数:8
相关论文
共 18 条
  • [1] BAYDUSH A, 1998, UNPUB RADIOLOGRAPHIC
  • [2] BAYESIAN IMAGE ESTIMATION OF DIGITAL CHEST RADIOGRAPHY - INTERDEPENDENCE OF NOISE, RESOLUTION, AND SCATTER FRACTION
    BAYDUSH, AH
    FLOYD, CE
    [J]. MEDICAL PHYSICS, 1995, 22 (08) : 1255 - 1261
  • [3] Spatially varying Bayesian image estimation
    Baydush, AH
    Floyd, CE
    [J]. ACADEMIC RADIOLOGY, 1996, 3 (02) : 129 - 136
  • [4] Improved Bayesian image estimation for digital chest radiography
    Baydush, AH
    Bowsher, JE
    Laading, JK
    Floyd, CE
    [J]. MEDICAL PHYSICS, 1997, 24 (04) : 539 - 545
  • [5] CHARACTERIZATION OF THE POINT SPREAD FUNCTION AND MODULATION TRANSFER-FUNCTION OF SCATTERED RADIATION USING A DIGITAL IMAGING-SYSTEM
    BOONE, JM
    ARNOLD, BA
    SEIBERT, JA
    [J]. MEDICAL PHYSICS, 1986, 13 (02) : 254 - 256
  • [6] Quality control phantom for digital chest radiography
    Chotas, HG
    Floyd, CE
    Johnson, GA
    Ravin, CE
    [J]. RADIOLOGY, 1997, 202 (01) : 111 - 116
  • [7] Dempster A. P., 1977, JRSSSB, V39, P1
  • [8] SINGLE-EXPOSURE DUAL-ENERGY COMPUTED RADIOGRAPHY - IMPROVED DETECTION AND PROCESSING
    ERGUN, DL
    MISTRETTA, CA
    BROWN, DE
    BYSTRIANYK, RT
    SZE, WK
    KELCZ, F
    NAIDICH, DP
    [J]. RADIOLOGY, 1990, 174 (01) : 243 - 249
  • [9] BAYESIAN RESTORATION OF CHEST RADIOGRAPHS - SCATTER COMPENSATION WITH IMPROVED SIGNAL-TO-NOISE RATIO
    FLOYD, CE
    BAYDUSH, AH
    LO, JY
    BOWSHER, JE
    RAVIN, CE
    [J]. INVESTIGATIVE RADIOLOGY, 1994, 29 (10) : 904 - 910
  • [10] QUANTITATIVE SCATTER MEASUREMENT IN DIGITAL RADIOGRAPHY USING A PHOTOSTIMULABLE PHOSPHOR IMAGING-SYSTEM
    FLOYD, CE
    LO, JY
    CHOTAS, HG
    RAVIN, CE
    [J]. MEDICAL PHYSICS, 1991, 18 (03) : 408 - 413