A compact electric potential sensor array for the acquisition and reconstruction of the 7-lead electrocardiogram without electrical charge contact with the skin

被引:23
作者
Harland, CJ [1 ]
Clark, TD
Peters, NS
Everitt, MJ
Stiffell, PB
机构
[1] Univ Sussex, Sch Sci & Technol, Ctr Phys Elect & Quantum Technol, Brighton BN1 9QT, E Sussex, England
[2] Imperial Coll Sch Med, London W2 1NY, England
关键词
electrocardiogram; electric potential sensor; cardiac sensor array;
D O I
10.1088/0967-3334/26/6/005
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Conventional electrocardiogram (ECG) systems make use of separate electrical connections to the arms and legs. These use a 'long baseline' for the voltage reference potential which in the case of precordial ECG leads is provided using a Wilson central terminal (WCT) wiring configuration. The aims of this project were (a) to construct compact, non-invasive surface ECG sensor arrays which would operate without the need for a WCT reference, (b) to obtain high quality precordial ECGs showing fine differences in ECG detail between small adjacent areas of the chest and (c) to reconstruct, from a compact array of four sensors, ECGs which closely match to the conventional 7-lead ECG system, but without the need for multiple wires and long baselines. In this paper, we describe two sensor array configurations which have been constructed using electric potential sensors (EPSs). We show high quality precordial ECGs obtained from small areas of the surface of the chest and show the different angular vectors (leads) in the frontal cardiac plane constructed using signals from the array elements. We suggest that these ECG arrays, which are simple to apply, should prove to be a valuable tool in providing useful information about the state of the heart.
引用
收藏
页码:939 / 950
页数:12
相关论文
共 12 条
[1]  
*EASI, 12 LEAD SYST
[2]   APPRAISAL OF A LOW-NOISE ELECTROCARDIOGRAM [J].
ELSHERIF, N ;
MEHRA, R ;
GOMES, JAC ;
KELEN, G .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1983, 1 (02) :456-467
[3]   High resolution ambulatory electrocardiographic monitoring using wrist-mounted electric potential sensors [J].
Harland, CJ ;
Clark, TD ;
Prance, RJ .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2003, 14 (07) :923-928
[4]   Remote detection of human electroencephalograms using ultrahigh input impedance electric potential sensors [J].
Harland, CJ ;
Clark, TD ;
Prance, RJ .
APPLIED PHYSICS LETTERS, 2002, 81 (17) :3284-3286
[5]   Electric potential probes - new directions in the remote sensing of the human body [J].
Harland, CJ ;
Clark, TD ;
Prance, RJ .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2002, 13 (02) :163-169
[6]  
HYTTINEN JAK, 1995, ELECTROENCEPHALOGRAP, P313
[7]  
MALMIVUO J, 2005, ELECTROENCEPHALOGRAP, P277
[8]   Assessment of derived 12-lead electrocardiograms using general and patient-specific reconstruction strategies at rest and during transient myocardial ischemia [J].
Nelwan, SP ;
Crater, SW ;
Green, CL ;
Johanson, P ;
van Dam, TB ;
Meij, SH ;
Simoons, ML ;
Krucoff, MW .
AMERICAN JOURNAL OF CARDIOLOGY, 2004, 94 (12) :1529-1533
[9]   Minimal lead sets for reconstruction of 12-lead electrocardiograms [J].
Nelwan, SP ;
Kors, JA ;
Meij, SH .
JOURNAL OF ELECTROCARDIOLOGY, 2000, 33 :163-166
[10]   Non-contact VLSI imaging using a scanning electric potential microscope [J].
Prance, RJ ;
Clark, TD ;
Prance, H ;
Clippingdale, A .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1998, 9 (08) :1229-1235