Amperometric monitoring of lactate accumulation in rabbit ischemic myocardium

被引:36
作者
Marzouk, SAM
Cosofret, VV
Buck, RP
Yang, H
Cascio, WE
Hassan, SSM
机构
[1] UNIV N CAROLINA,DEPT CHEM,CHAPEL HILL,NC 27599
[2] UNIV N CAROLINA,DEPT MED,CHAPEL HILL,NC 27599
[3] AIN SHAMS UNIV,DEPT CHEM,CAIRO,EGYPT
基金
美国国家卫生研究院;
关键词
amperometric biosensors; ischemia; lactate monitoring; papillary muscle;
D O I
10.1016/S0039-9140(96)02175-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Fabrication and characterization of miniature, flexible, planar biosensors for monitoring L-lactate accumulation in an ischemic myocardium are described. Three configurations of Au-based electrodes were fabricated by a photolithographic technique on flexible polyimide Kapton(R) foil. All sensors are based on an immobilized lactate oxidase with amperometric detection of the enzymatically produced hydrogen peroxide al a platinum-electroplated-gold base electrode polarized at 0.5 V versus Ag/AgCl. An inner electropolymeric layer is used to prevent electrode fouling and to reject the interference effects of easily oxidizable molecules. In addition, a diffusion controlling outer layer that greatly enhances the linear dynamic range of the sensor, is obtained by casting a polyurethane external film. The developed sensor was evaluated in vitro and proved to have high selectivity, good operational stability, good accuracy and precision (average recovery = 102.3 +/- 0.4% for control sera), fast response time (t(95) = 20 s) and high upper limit of the linear dynamic range (25-80 mM, with sensitivity of 1.7-0.4 nA mM(-1), respectively at PO2 = 15 mmHg). Subsequently, the sensor was brought into direct contact with the surface of the rabbit papillary muscle and used for continuous quantitative monitoring of extracellular lactate accumulation during no-flow ischemia. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:1527 / 1541
页数:15
相关论文
共 45 条
[1]   A CONTINUOUS, IMPLANTABLE LACTATE SENSOR [J].
BAKER, DA ;
GOUGH, DA .
ANALYTICAL CHEMISTRY, 1995, 67 (09) :1536-1540
[2]   A RELIABLE L-LACTATE ELECTRODE WITH A NEW MEMBRANE FOR ENZYME IMMOBILIZATION FOR AMPEROMETRIC ASSAY OF LACTATE [J].
BARDELETTI, G ;
SECHAUD, F ;
COULET, PR .
ANALYTICA CHIMICA ACTA, 1986, 187 :47-54
[3]  
BLUM W, 1949, PRINCIPLES ELECTROPL, P382
[4]   MICROFABRICATED SENSOR ARRAYS SENSITIVE TO PH AND K+ FOR IONIC DISTRIBUTION MEASUREMENTS IN THE BEATING HEART [J].
COSOFRET, VV ;
ERDOSY, M ;
JOHNSON, TA ;
BUCK, RP ;
ASH, RB ;
NEUMAN, MR .
ANALYTICAL CHEMISTRY, 1995, 67 (10) :1647-1653
[5]   ENDOCARDIAL AND EPICARDIAL INTERSTITIAL PURINES AND LACTATE DURING GRADED ISCHEMIA [J].
DELYANI, JA ;
VANWYLEN, DGL .
AMERICAN JOURNAL OF PHYSIOLOGY, 1994, 266 (03) :H1019-H1026
[6]   ELECTROPOLYMERIZED O-PHENYLENEDIAMINE FILM AS MEANS OF IMMOBILISING LACTATE OXIDASE FOR A L-LACTATE BIOSENSOR [J].
DEMPSEY, E ;
WANG, J ;
SMYTH, MR .
TALANTA, 1993, 40 (03) :445-451
[7]   ELECTROPOLYMERIZED FILMS TO PREVENT INTERFERENCES AND ELECTRODE FOULING IN BIOSENSORS [J].
GEISE, RJ ;
ADAMS, JM ;
BARONE, NJ ;
YACYNYCH, AM .
BIOSENSORS & BIOELECTRONICS, 1991, 6 (02) :151-160
[8]   IMMOBILIZATION OF LACTATE OXIDASE IN A POLY(VINYL ALCOHOL) MATRIX ON PLATINIZED GRAPHITE-ELECTRODES BY CHEMICAL CROSS-LINKING WITH ISOCYANATE [J].
HAJIZADEH, K ;
HALSALL, HB ;
HEINEMAN, WR .
TALANTA, 1991, 38 (01) :37-47
[9]   EXCITATORY FACTORS IN VENTRICULAR TACHYCARDIA RESULTING FROM MYOCARDIAL ISCHEMIA - POTASSIUM A MAJOR EXCITANT [J].
HARRIS, AS ;
BISTENI, A ;
RUSSELL, RA ;
BRIGHAM, JC ;
FIRESTONE, JE .
SCIENCE, 1954, 119 (3085) :200-203
[10]   DIRECT MEASUREMENT OF GLUTAMATE RELEASE IN THE BRAIN USING A DUAL ENZYME-BASED ELECTROCHEMICAL SENSOR [J].
HU, YB ;
MITCHELL, KM ;
ALBAHADILY, FN ;
MICHAELIS, EK ;
WILSON, GS .
BRAIN RESEARCH, 1994, 659 (1-2) :117-125