Inappropriate tachycardia detection by a biventricular Implantable cardioverter defibrillator

被引:18
作者
Garcia-Moran, E [1 ]
Mont, L [1 ]
Brugada, J [1 ]
机构
[1] Univ Barcelona, Hosp Clin, Cardiovasc Inst, Arrhythmia Sect, E-08036 Barcelona, Spain
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2002年 / 25卷 / 01期
关键词
implantable cardioverter defibrillator; pacing; oversensing; ventricular tachycardia;
D O I
10.1046/j.1460-9592.2002.00123.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This case describes "double counting" of intrinsic conduction due to the combination of intraventricular conduction delay and the merging of the electrograms from the right and left ventricle (coronary sinus) leads in a biventricular cardioverter defibrillator. This situation prompted asymptomatic antitachycardia pacing. Lengthening of AV conduction by betablocker solved the problem.
引用
收藏
页码:123 / 124
页数:2