Improving IVUS Palpography by Incorporation of Motion Compensation Based on Block Matching and Optical Flow

被引:12
作者
Danilouchkine, Mikhail G. [1 ]
Mastik, Frits [1 ]
van der Steen, Antonius F. W. [1 ,2 ]
机构
[1] Erasmus MC, Dept Biomed Engn, Ctr Thorax, Rotterdam, Netherlands
[2] Interuniv Cardiol Inst Netherlands, Utrecht, Netherlands
关键词
D O I
10.1109/TUFFC.947
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Intravascular ultrasound (IVUS) strain imaging of the luminal layer in coronary arteries, coined as IVUS palpography, utilizes conventional radio frequency (RF) Signals acquired kit 2 different levels of a compressional load. The signals are cross-correlated to obtain the microscopic tissue displacements, which call be directly translated into local strain of the vessel wall. However, (apparent) tissue motion and nonuniform deformation of the vessel wall, due to catheter wiggling, reduce signal correlation and result in invalid strain estimates. Implications of probe motion were studied oil the tissue-mimicking phantom. The measured circumferential tissue displacement and level of the speckle decorrelation amounted to 12 degrees and 0.58, respectively, for the catheter displacement of 456 mu m. To compensate for the motion artifacts in IVUS palpography, a novel method based on the feature-based scale-space optical flow (OF), and classical block matching (BM) algorithm, were employed. The computed OF vector and BM displacement fields quantify the amount of local tissue misalignment in consecutive frames. Subsequently, the extracted circumferential displacements are used to realign the signals before strain computation. Motion compensation reduces the RF signal decorrelation and increases the number of valid strain estimates. The advantage of applying the motion correction in IVUS palpography was demonstrated in a midscale validation study on 14 in vivo pullbacks. Both methods substantially increase the number of valid strain estimates in the partial and compounded palpograms. Mean relative improvement in the number of valid strain estimates with motion compensation in comparison to one without motion compensation amounts to 28% and 14%, respectively. Implementation of motion compensation methods boosts the diagnostic value of IVUS palpography
引用
收藏
页码:2392 / 2404
页数:13
相关论文
共 71 条
[1]   Model-based reconstructive elasticity imaging of deep venous thrombosis [J].
Aglyamov, S ;
Skovoroda, AR ;
Rubin, JM ;
O'Donnell, M ;
Emelianov, SY .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (05) :521-531
[2]   Reduction of signal decorrelation from mechanical compression of tissues by temporal stretching: Applications to elastography [J].
Alam, SK ;
Ophir, J .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (01) :95-105
[3]   An adaptive strain estimator for elastography [J].
Alam, SK ;
Ophir, J ;
Konofagou, EE .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (02) :461-472
[4]   Axial movement of the intravascular ultrasound probe during the cardiac cycle: Implications for three-dimensional reconstruction and measurements of coronary dimensions [J].
Arbab-Zadeh, A ;
DeMaria, AN ;
Penny, WF ;
Russo, RJ ;
Kimura, BJ ;
Bhargava, V .
AMERICAN HEART JOURNAL, 1999, 138 (05) :865-872
[5]   Young's modulus reconstruction of vulnerable atherosclerotic plaque components using deformable curves [J].
Baldewsing, RA ;
Mastik, F ;
Schaar, JA ;
Serruys, PW ;
van der Steen, AFW .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2006, 32 (02) :201-210
[6]   An inverse method for imaging the local elasticity of atherosclerotic coronary plaques [J].
Baldewsing, Radjkumarsing Radj A. ;
Danilouchkine, Mikhail G. ;
Mastik, Frits ;
Schaar, Johannes A. ;
Serruys, Patrick W. ;
van der Steen, Antonius F. W. .
IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2008, 12 (03) :277-289
[7]   Evaluation of differential optical flow techniques on synthesized echo images [J].
Baraldi, P ;
Sarti, A ;
Lamberti, C ;
Prandini, A ;
Sgallari, F .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (03) :259-272
[8]   The computation of optical flow [J].
Beauchemin, SS ;
Barron, JL .
ACM COMPUTING SURVEYS, 1995, 27 (03) :433-467
[9]   Improving motion estimation by accounting for local image distortion [J].
Behar, V ;
Adam, D ;
Lysyansky, P ;
Friedman, Z .
ULTRASONICS, 2004, 43 (01) :57-65
[10]   The robust estimation of multiple motions: Parametric and piecewise-smooth flow fields [J].
Black, MJ ;
Anandan, P .
COMPUTER VISION AND IMAGE UNDERSTANDING, 1996, 63 (01) :75-104