Echo planar MRI of the heart on a standard system: Validation of measurements of left ventricular function and mass

被引:72
作者
Lamb, HJ [1 ]
Doornbos, J [1 ]
vanderVelde, EA [1 ]
Kruit, MC [1 ]
Reiber, JHC [1 ]
deRoos, A [1 ]
机构
[1] LEIDEN UNIV HOSP,DEPT MED STAT,NL-2333 AA LEIDEN,NETHERLANDS
关键词
magnetic resonance imaging; heart; ventricles; techniques;
D O I
10.1097/00004728-199611000-00014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: Our goal was to validate cardiac measurements derived from multishot echo planar MRI (EPI) as compared with the well validated conventional GRE technique. Method: Ten healthy subjects underwent breath-hold EPI and non-breath-hold GRE imaging in the short axis orientation of the left ventricle (LV) on a standard 1.5 T MR system. Ten section levels were obtained to encompass the entire LV. Measurements were obtained of end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF), LV mass, time to end-systole (TES), and peak filling rate (PFR). Analysis of variance was performed to determine agreement between GRE- and EPI-derived measurements. Results: The acquisition time for EPI was much shorter than that for GRE (2.5 vs. 15 min). Both imaging techniques yielded good quality images allowing LV volumetrics. Agreement between CRE and EPI was best for measurements of EDV, SV, and LV mass; somewhat less agreement was found for ESV, EF, TES, and PFR. The intraobserver variability for measuring TES and PFR was higher for GRE than EPI (one sided F test; critical values at p = 0.05 were >3.18). Conclusion: Multishot EPI of the heart provides accurate measurements of LV function and mass in a time-efficient manner.
引用
收藏
页码:942 / 949
页数:8
相关论文
共 27 条
  • [1] AUTOMATED MYOCARDIAL EDGE-DETECTION FROM BREATH-HOLD CINE-MR IMAGES - EVALUATION OF LEFT-VENTRICULAR VOLUMES AND MASS
    BALDY, C
    DOUEK, P
    CROISILLE, P
    MAGNIN, IE
    REVEL, D
    AMIEL, M
    [J]. MAGNETIC RESONANCE IMAGING, 1994, 12 (04) : 589 - 598
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] INFLUENCE OF TEMPORAL SMOOTHING ON QUANTITATION OF LEFT-VENTRICULAR FUNCTION BY GATED BLOOD POOL SCINTIGRAPHY
    BONOW, RO
    BACHARACH, SL
    CRAWFORDGREEN, C
    GREEN, MV
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1989, 64 (14) : 921 - 925
  • [4] INTERLEAVED ECHO-PLANAR IMAGING ON A STANDARD MRI SYSTEM
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    THOMPSON, RM
    JACK, CR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (01) : 67 - 72
  • [5] ECHO-PLANAR IMAGING OF THE LIVER WITH A STANDARD MR-IMAGING SYSTEM
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    FELMLEE, JP
    GRIMM, RC
    [J]. RADIOLOGY, 1993, 189 (01) : 259 - 264
  • [6] DETERMINATION OF LEFT-VENTRICULAR VOLUME AND MASS WITH USE OF BIPHASIC SPIN-ECHO MR IMAGING - COMPARISON WITH CINE MR
    CAPUTO, GR
    SUZUKI, JI
    KONDO, C
    CHO, H
    QUAIFE, RA
    HIGGINS, CB
    PARKER, DL
    [J]. RADIOLOGY, 1990, 177 (03) : 773 - 777
  • [7] NORMAL HEART - EVALUATION WITH ECHO-PLANAR MR-IMAGING
    DAVIS, CP
    MCKINNON, GC
    DEBATIN, JF
    WETTER, D
    EICHENBERGER, AC
    DUEWELL, S
    VONSCHULTHESS, GK
    [J]. RADIOLOGY, 1994, 191 (03) : 691 - 696
  • [8] SINGLE-SHOT VERSUS INTERLEAVED ECHO-PLANAR MR-IMAGING - APPLICATION TO VISUALIZATION OF CARDIAC-VALVE LEAFLETS
    DAVIS, CP
    MCKINNON, GC
    DEBATIN, JF
    DUEWELL, S
    VONSCHULTHESS, GK
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (01): : 107 - 112
  • [9] SEGMENTED TURBOFLASH - METHOD FOR BREATH-HOLD MR IMAGING OF THE LIVER WITH FLEXIBLE CONTRAST
    EDELMAN, RR
    WALLNER, B
    SINGER, A
    ATKINSON, DJ
    SAINI, S
    [J]. RADIOLOGY, 1990, 177 (02) : 515 - 521
  • [10] ECHO-PLANAR MR-IMAGING
    EDELMAN, RR
    WIELOPOLSKI, P
    SCHMITT, F
    [J]. RADIOLOGY, 1994, 192 (03) : 600 - 612