Desorption electrospray ionisation (DESI) for beginners - how to adjust settings for tissue imaging

被引:26
作者
Bodzon-Kulakowska, Anna [1 ]
Drabik, Anna [1 ]
Ner, Joanna [1 ]
Kotlinska, Jolanta Helena [2 ]
Suder, Piotr [1 ]
机构
[1] AGH Univ Sci & Technol, Fac Mat Sci & Ceram, Dept Biochem & Neurobiol, PL-30059 Krakow, Poland
[2] Med Univ Lublin, Dept Pharmacol & Pharmacodynam, PL-20093 Lublin, Poland
关键词
AUTOMATIC GAIN-CONTROL; MASS-SPECTROMETRY;
D O I
10.1002/rcm.6755
中图分类号
Q5 [生物化学];
学科分类号
070307 [化学生物学];
摘要
RATIONALEDesorption electrospray ionisation (DESI) is the ambient technique used for surface imaging. Despite its simplicity, the proper use of this technique is not easy, and usually leads to discouragement, especially in the case of biological sample measurements. Here, we present some tips and tricks which may be helpful during a complex process of ion source optimisation to achieve the desired results. METHODSRat liver tissue as an example of a biological sample and a surface covered with rhodamine-containing marker were measured using a DESI ion source (OMNIspray source, Prosolia, Indianapolis, IN, USA) connected to a AmaZon ETD ion trap mass spectrometer (Bruker Daltonics, Bremen, Germany). RESULTSThe geometry of the ion source (nebulisation capillary angle, its distance to the surface, and to the MS inlet), and other settings like nebulising gas pressure, solvent flow and capillary voltage, were changed during the optimisation process. The results obtained for different parameters are presented. CONCLUSIONSDifferences between the results and the method of optimisation for biological and non-biological samples were shown. The influence of different parameters on the quality of mass spectra was indicated. Optimal parameters for the tissue and non-biological sample analysis were suggested. Copyright (c) 2013 John Wiley & Sons, Ltd.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 12 条
[1]
Simulated splashes: Elucidating the mechanism of desorption electrospray ionization mass spectrometry [J].
Costa, Anthony B. ;
Cooks, R. Graham .
CHEMICAL PHYSICS LETTERS, 2008, 464 (1-3) :1-8
[2]
Deconstructing Desorption Electrospray Ionization: Independent Optimization of Desorption and Ionization by Spray Desorption Collection [J].
Douglass, Kevin A. ;
Jain, Shashank ;
Brandt, William R. ;
Venter, Andre R. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2012, 23 (11) :1896-1902
[3]
Desorption electrospray ionization mass spectrometry for lipid characterization and biological tissue imaging [J].
Eberlin, Livia S. ;
Ferreira, Christina R. ;
Dill, Allison L. ;
Ifa, Demian R. ;
Cooks, R. Graham .
BIOCHIMICA ET BIOPHYSICA ACTA-MOLECULAR AND CELL BIOLOGY OF LIPIDS, 2011, 1811 (11) :946-960
[4]
Desorption Electrospray Ionization Imaging Mass Spectrometry of Lipids in Rat Spinal Cord [J].
Girod, Marion ;
Shi, Yunzhou ;
Cheng, Ji-Xin ;
Cooks, R. Graham .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2010, 21 (07) :1177-1189
[5]
Calibration Function for the Orbitrap FTMS Accounting for the Space Charge Effect [J].
Gorshkov, Mikhail V. ;
Good, David M. ;
Lyutvinskiy, Yaroslav ;
Yang, Hongqian ;
Zubarev, Roman A. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2010, 21 (11) :1846-1851
[6]
A Dual Pressure Linear Ion Trap Orbitrap Instrument with Very High Sequencing Speed [J].
Olsen, Jesper V. ;
Schwartz, Jae C. ;
Griep-Raming, Jens ;
Nielsen, Michael L. ;
Damoc, Eugen ;
Denisov, Eduard ;
Lange, Oliver ;
Remes, Philip ;
Taylor, Dennis ;
Splendore, Maurizio ;
Wouters, Eloy R. ;
Senko, Michael ;
Makarov, Alexander ;
Mann, Matthias ;
Horning, Stevan .
MOLECULAR & CELLULAR PROTEOMICS, 2009, 8 (12) :2759-2769
[7]
Automatic gain control in mass spectrometry using a jet disrupter electrode in an electrodynamic ion funnel [J].
Page, JS ;
Bogdanov, B ;
Vilkov, AN ;
Prior, DC ;
Buschbach, MA ;
Tang, K ;
Smith, RD .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2005, 16 (02) :244-253
[8]
Ross M. H., 2006, HISTOLOGY TEXT ATLAS, P532
[9]
Mass spectrometry sampling under ambient conditions with desorption electrospray ionization [J].
Takáts, Z ;
Wiseman, JM ;
Gologan, B ;
Cooks, RG .
SCIENCE, 2004, 306 (5695) :471-473
[10]
Droplet dynamics and ionization mechanisms in desorption electrospray ionization mass spectrometry [J].
Venter, Andre ;
Sojka, Paul E. ;
Cooks, R. Graham .
ANALYTICAL CHEMISTRY, 2006, 78 (24) :8549-8555