The HemoCue®, a point of care B-hemoglobin photometer, measures hemoglobin concentrations accurately when mixed in vitro with canine plasma and three hemoglobin-based oxygen carriers (HBOC)

被引:22
作者
Jahr, JS
Lurie, F
Driessen, B
Davis, JA
Gosselin, R
Gunther, RA
机构
[1] Univ Calif Los Angeles, Hlth Sci Ctr, Dept Anesthesiol, Sch Med, Los Angeles, CA 90095 USA
[2] Charles R Drew Univ Med & Sci, Martin Luther King Jr Drew Med Ctr, Los Angeles, CA 90059 USA
[3] Univ Penn, New Bolton Ctr, Dept Clin Studies, Kennett Sq, PA 19348 USA
[4] Univ Calif Davis, Sch Med, Dept Anesthesiol & Pain Med, Sacramento, CA 95817 USA
[5] Univ Calif Davis, Sch Med, Dept Surg, Sacramento, CA 95817 USA
[6] Univ Calif Davis, Sch Med, Dept Med Pathol, Sacramento, CA 95817 USA
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 2002年 / 49卷 / 03期
关键词
D O I
10.1007/BF03020522
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Purpose: Accuracy of measurement of low hemoglobin concentrations using the HemoCue(R), a B-hemoglobin photometer (HemoCue AB, Angelholm, Sweden) may exhibit significant variability. Infusion of hemoglobin-based oxygen carriers (HBOC) result in low concentrations of plasma hemoglobin. Our study assessed B-hemoglobin photometer measurement accuracy of three HBOC: (hemoglobin glutamer-200 (bovine; Oxyglobin(R), Biopure Corp., Cambridge, MA,USA); hemoglobin glutamer-250 (bovine; Hemopure(R), Biopure Corp, Cambridge, MA, USA), and hemoglobin-raffimer, (human; Hemolink(TM), Hemosol, Inc., Toronto, Ontario, Canada). Methods: In the laboratory, 45 split canine plasma samples were mixed with hemoglobin glutamer-200 (8.13, 16.25, 32.5 g(.)L(-1) concentrations), 45 samples were mixed with hemoglobin glutamer-250 (8.13, 16.25, 32.5 g(.)L(-1) concentrations), 45 with hemoglobin-raffimer (12.5, 25.0, 50.0 g(.)L(-1) concentrations), and measured. Plasma samples without HBOC served as control. Hemoglobin concentration was determined by a laboratory analyzer (Coulter Corporation, Hiafeah, FL, USA) and B-hemoglobin photometer (HemoCue(R), Angelholm, Sweden). Two independent technicians performed blinded sample measurements and randomly tested each sample five times. Results were analyzed according to Bland and Altman analysis. Results: B-hemoglobin photometer demonstrated high repeatability for all three HBOCs. Repeatability coefficients were 0,37 g(.)L(-1) and 0.48 g(.)L(-1) for hemoglobin glutamer-200, 0.39 g(.)L(-1) and 0.4 g(.)L(-1) for hemoglobin glutamer-250 and 1.07 g(.)L(-1) and 0.85 g(.)L(-1) for hemoglobin-raffimer. An acceptable agreement was found between the B-hemoglobin photometer and the laboratory analyzer for all three HBOCs tested. Conclusion: The B-hemoglobin photometer accurately determined the concentration of three HBOC solutions dissolved in canine plasma.
引用
收藏
页码:243 / 248
页数:6
相关论文
共 12 条
[1]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[2]   Perioperative on-site haemoglobin determination: As accurate as laboratory values? [J].
Jaeger, M ;
Ashbury, T ;
Adams, M ;
Duncan, P .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1996, 43 (08) :795-798
[3]  
Jahr J S, 2000, Clin Lab Sci, V13, P210
[4]  
Jahr J S, 2000, Clin Lab Sci, V13, P173
[5]   Oxygen saturation measurements in canine blood containing hemoglobin glutamer-200 (bovine): In vitro validation of the NOVA CO-Oximeter [J].
Jahr, JS ;
Driessen, B ;
Lurie, F ;
Tang, ZP ;
Louie, RF ;
Kost, G .
VETERINARY CLINICAL PATHOLOGY, 2001, 30 (01) :39-45
[6]   A novel approach to measuring circulating blood volume: The use of a hemoglobin-based oxygen carrier in a rabbit model [J].
Jahr, JS ;
Lurie, F ;
Xi, S ;
Golkaryeh, M ;
Kuznetsova, O ;
Kullar, R ;
Driessen, B .
ANESTHESIA AND ANALGESIA, 2001, 92 (03) :609-614
[7]  
Kasper SM, 1996, ANESTH ANALG, V83, P921
[8]   Evaluation of the HemoCue® for measuring intra-operative haemoglobin concentrations:: a comparison with the Coulter Max-M® [J].
Lardi, AM ;
Hirst, C ;
Mortimer, AJ ;
McCollum, CN .
ANAESTHESIA, 1998, 53 (04) :349-352
[9]   Bovine hemoglobin-based oxygen carrier (HBOC-201) for resuscitation of uncontrolled, exsanguinating liver injury in swine [J].
Manning, JE ;
Katz, LM ;
Brownstein, MR ;
Pearce, LB ;
Gawryl, MS ;
Baker, CC .
SHOCK, 2000, 13 (02) :152-159
[10]   Comparing methods of clinical measurement: Reporting standards for Bland and Altman analysis [J].
Mantha, S ;
Roizen, MF ;
Fleisher, LA ;
Thisted, R ;
Foss, J .
ANESTHESIA AND ANALGESIA, 2000, 90 (03) :593-602