Computational and experimental evaluation of the fluid dynamics and hemocompatibility of the CentriMag blood pump

被引:133
作者
Zhang, JT
Gellman, B
Koert, A
Dasse, KA
Gilbert, RJ
Griffith, BP
Wu, ZJJ
机构
[1] Univ Maryland, Sch Med, Dept Surg, Artificial Organs Lab, Baltimore, MD 21201 USA
[2] Levitronix LLC, Waltham, MA USA
[3] MIT, Dept Mech Engn, Cambridge, MA 02139 USA
关键词
computational fluid dynamics; hemolysis modeling; magnetically levitated rotor; CentriMag blood pump; ventricular assist device;
D O I
10.1111/j.1525-1594.2006.00203.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The CentriMag centrifugal blood pump is a newly developed ventricular assist device based on magnetically levitated bearingless rotor technology. A combined computational and experimental study was conducted to characterize the hemodynamic and hemocompatibility performances of this novel blood pump. Both the three-dimensional flow features of the CentriMag blood pump and its hemolytic characteristics were analyzed using computational fluid dynamics (CFD)-based modeling. The hydraulic pump performance and hemolysis level were quantified experimentally. The CFD simulation demonstrated a clean and streamlined flow field in the main components of the CentriMag blood pump. The predicted results by hemolysis model indicated no significant high shear stress regions in the pump. A comparison of CFD predictions and experimental results showed good agreements. The relatively large gap passages (1.5 mm) between the outer rotor walls and the lower housing cavity walls provide a very good surface washing through a secondary flow path while the shear stresses in the secondary flow paths are reduced, resulting in a low rate of hemolysis ([Normalized Index of Hemolysis] NIH = 0.0029 +/- 0.006) without a decrease of the pump's hydrodynamic performance (pressure head: 352 mm Hg at a flow rate of 5.0 L/min and a rotational speed of 4000 rpm).
引用
收藏
页码:168 / 177
页数:10
相关论文
共 16 条
[1]  
[Anonymous], 1972, BIOMECHANICS
[2]  
[Anonymous], F184197 ASTM
[3]  
[Anonymous], F183097 ASTM
[4]   3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :590-596
[5]   Computational fluid dynamics analysis of a Maglev centrifugal left ventricular assist device [J].
Burgreen, GW ;
Loree, HM ;
Bourque, K ;
Dague, C ;
Poirier, VL ;
Farrar, D ;
Hampton, E ;
Wu, ZJ ;
Gempp, TM ;
Schöb, R .
ARTIFICIAL ORGANS, 2004, 28 (10) :874-880
[6]   A prototype HeartQuest ventricular assist device for particle image velocimetry measurements [J].
Day, SW ;
McDaniel, JC ;
Wood, HG ;
Allaire, PE ;
Song, XW ;
Lemire, PP ;
Miles, SD .
ARTIFICIAL ORGANS, 2002, 26 (11) :1002-1005
[7]  
Fung Y., 1993, BIOMECHANICS MECH PR
[8]   Fast three-dimensional numerical hemolysis approximation [J].
Garon, A ;
Farinas, MI .
ARTIFICIAL ORGANS, 2004, 28 (11) :1016-1025
[9]   ESTIMATION OF SHEAR STRESS-RELATED BLOOD DAMAGE IN HEART-VALVE PROSTHESES - INVITRO COMPARISON OF 25 AORTIC VALVES [J].
GIERSIEPEN, M ;
WURZINGER, LJ ;
OPITZ, R ;
REUL, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1990, 13 (05) :300-306
[10]   Evaluation of hemolysis in the VentrAssist implantable rotary blood pump [J].
James, NL ;
Wilkinson, CM ;
Lingard, NL ;
van der Meer, AL ;
Woodard, JC .
ARTIFICIAL ORGANS, 2003, 27 (01) :108-113