Capillary electrophoresis tandem mass spectrometry with array detection

被引:1
作者
Deterding, LJ
Tomer, KB
Wellemans, JMY
Cerny, RL
Gross, ML
机构
[1] NIEHS, Struct Biol Lab, Res Triangle Pk, NC 27709 USA
[2] Univ Nebraska, Dept Chem, Lincoln, NE 68588 USA
来源
EUROPEAN MASS SPECTROMETRY | 1999年 / 5卷 / 01期
关键词
capillary electrophoresis; tandem mass spectrometry; array detection; coaxial continuous flow FAB; peptides;
D O I
10.1255/ejms.247
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Mixtures of standard compounds of biological interest were analyzed by capillary electrophoresis in conjunction with tandem four-sector mass spectrometry (MS/MS) using an array detector. Capillary electrophoresis offers extremely high separation efficiencies while array detection allows for the simultaneous acquisition of a large fraction of a tandem mass spectrum and high utilization of the sample ions produced. Consequently, improvements in sensitivity and structural information were observed for the combined techniques in comparison to similar measurements with a four-sector and a single-point detector. Coaxial continuous-flow fast atom bombardment was used to create (M + H)(+) precursor ions of the separated analytes and their product ion spectra were acquired from femtomole levels in electrophoretic real-time, The limit-of-detection for the acquisition of MS/MS data from electrophoretically-separated peptides was 35-100 fmol, The MS/MS data acquired at these levels showed cleavages of the peptide backbone (i.e. to give a and y ions) and production of immonium ions from the constituent amino acids. Acquisition of the spectra via the array detector is more routine acid, at the 100 femtomole level, provides more structurally informative ions and yields better ion statistics than comparable spectra acquired with a point detector.
引用
收藏
页码:33 / 40
页数:8
相关论文
共 32 条
[1]   CONTRIBUTIONS OF MASS-SPECTROMETRY TO PEPTIDE AND PROTEIN-STRUCTURE [J].
BIEMANN, K .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1988, 16 (1-12) :99-111
[2]  
Burlingame AL, 1996, ANAL CHEM, V68, pR599, DOI 10.1021/a1960021u
[3]  
CAPRIOLI RM, 1990, CONTINUOUS-FLOW FAST ATOM BOMBARDMENT MASS SPECTROMETRY, P93
[4]  
De Wit JSM, 1988, RAPID COMMUN MASS SP, V2, P100
[5]   COAXIAL CONTINUOUS-FLOW FAST ATOM BOMBARDMENT IN CONJUNCTION WITH TANDEM MASS-SPECTROMETRY FOR THE ANALYSIS OF BIOMOLECULES [J].
DETERDING, LJ ;
MOSELEY, MA ;
TOMER, KB ;
JORGENSON, JW .
ANALYTICAL CHEMISTRY, 1989, 61 (22) :2504-2511
[6]   MICRO-ELECTROSPRAY MASS-SPECTROMETRY - ULTRA-HIGH-SENSITIVITY ANALYSIS OF PEPTIDES AND PROTEINS [J].
EMMETT, MR ;
CAPRIOLI, RM .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (07) :605-613
[7]   AN APPROACH TO CORRELATE TANDEM MASS-SPECTRAL DATA OF PEPTIDES WITH AMINO-ACID-SEQUENCES IN A PROTEIN DATABASE [J].
ENG, JK ;
MCCORMACK, AL ;
YATES, JR .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (11) :976-989
[8]   Protein identification by capillary zone electrophoresis/microelectrospray ionization-tandem mass spectrometry at the subfemtomole level [J].
Figeys, D ;
vanOostveen, I ;
Ducret, A ;
Aebersold, R .
ANALYTICAL CHEMISTRY, 1996, 68 (11) :1822-1828
[9]   Identification of proteins by capillary electrophoresis tandem mass spectrometry - Evaluation of an on-line solid-phase extraction device [J].
Figeys, D ;
Ducret, A ;
Aebersold, R .
JOURNAL OF CHROMATOGRAPHY A, 1997, 763 (1-2) :295-306
[10]   High sensitivity identification of proteins by electrospray ionization tandem mass spectrometry: Initial comparison between an ion trap mass spectrometer and a triple quadrupole mass spectrometer [J].
Figeys, D ;
Aebersold, R .
ELECTROPHORESIS, 1997, 18 (3-4) :360-368