High sensitivity identification of proteins by electrospray ionization tandem mass spectrometry: Initial comparison between an ion trap mass spectrometer and a triple quadrupole mass spectrometer

被引:71
作者
Figeys, D [1 ]
Aebersold, R [1 ]
机构
[1] UNIV WASHINGTON, DEPT MOL BIOTECHNOL, SEATTLE, WA 98195 USA
关键词
solid phase micro extraction; capillary zone electrophoresis; tandem mass spectrometry; microelectrospray; protein identification; two-dimensional polyacrylamide gel electrophoresis;
D O I
10.1002/elps.1150180310
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Recently, we have shown that a solid-phase-microextraction/capillary electrophoresis device coupled to an electrospray ionization triple quadrupole mass spectrometer through a microelectrospray interface represents a powerful analytical system for the rapid, conclusive and sensitive identification of proteins separated by gel electrophoresis. Here we report on the successful coupling of the same device to an electrospray ionization ion trap mass spectrometer and on the comparative evaluation of the performance of the triple quadrupole and ion-trap-based systems, In the ion trap mass spectrometer-based system, using a tryptic digest of a calibrated bovine serum albumin sample, we achieved limits of detection in the single mass spectrometry (MS) and tandem MS mode, respectively, of 400 amol (if 20 mu L of solution at a concentration of 20 amol/mu L was applied). The system mras also successfully used to identify six yeast proteins isolated from a single analytical two-dimensional polyacrylamide gel. For the detection of unfragmented peptide ions both systems showed comparable sensitivity, whereas the ion-trap-based system showed superior performance with fragment ion spectra.
引用
收藏
页码:360 / 368
页数:9
相关论文
共 31 条
[1]   MICRO-ELECTROSPRAY - ZEPTOMOLE-ATTOMOLE PER MICROLITER SENSITIVITY FOR PEPTIDES [J].
ANDREN, PE ;
EMMETT, MR ;
CAPRIOLI, RM .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (09) :867-869
[2]   THE USE OF SOLID-PHASE CONCENTRATORS FOR ONLINE PRECONCENTRATION OF METALLOTHIONEIN PRIOR TO ISOFORM SEPARATION BY CAPILLARY ZONE ELECTROPHORESIS [J].
BEATTIE, JH ;
SELF, R ;
RICHARDS, MP .
ELECTROPHORESIS, 1995, 16 (03) :322-328
[3]  
Bonfichi R, 1995, J MASS SPECTROM, pS95
[4]   RAPID MASS-SPECTROMETRIC PEPTIDE SEQUENCING AND MASS MATCHING FOR CHARACTERIZATION OF HUMAN-MELANOMA PROTEINS ISOLATED BY 2-DIMENSIONAL PAGE [J].
CLAUSER, KR ;
HALL, SC ;
SMITH, DM ;
WEBB, JW ;
ANDREWS, LE ;
TRAN, HM ;
EPSTEIN, LB ;
BURLINGAME, AL .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1995, 92 (11) :5072-5076
[5]   Characterization of human serum amyloid A protein isoforms separated by two-dimensional electrophoresis by liquid chromatography electrospray ionization tandem mass spectrometry [J].
Ducret, A ;
Bruun, CF ;
Bures, EJ ;
Marhaug, G ;
Husby, G ;
Aebersold, R .
ELECTROPHORESIS, 1996, 17 (05) :866-876
[6]   MICRO-ELECTROSPRAY MASS-SPECTROMETRY - ULTRA-HIGH-SENSITIVITY ANALYSIS OF PEPTIDES AND PROTEINS [J].
EMMETT, MR ;
CAPRIOLI, RM .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (07) :605-613
[7]   AN APPROACH TO CORRELATE TANDEM MASS-SPECTRAL DATA OF PEPTIDES WITH AMINO-ACID-SEQUENCES IN A PROTEIN DATABASE [J].
ENG, JK ;
MCCORMACK, AL ;
YATES, JR .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (11) :976-989
[8]   ELECTROSPRAY IONIZATION-PRINCIPLES AND PRACTICE [J].
FENN, JB ;
MANN, M ;
MENG, CK ;
WONG, SF ;
WHITEHOUSE, CM .
MASS SPECTROMETRY REVIEWS, 1990, 9 (01) :37-70
[9]   Protein identification by capillary zone electrophoresis/microelectrospray ionization-tandem mass spectrometry at the subfemtomole level [J].
Figeys, D ;
vanOostveen, I ;
Ducret, A ;
Aebersold, R .
ANALYTICAL CHEMISTRY, 1996, 68 (11) :1822-1828
[10]   Protein identification by solid phase microextraction-capillary zone electrophoresis-microelectrospray-tandem mass spectrometry [J].
Figeys, D ;
Ducret, A ;
Yates, JR ;
Aebersold, R .
NATURE BIOTECHNOLOGY, 1996, 14 (11) :1579-1583