Effect of polyethylene wear on osteolysis in cementless primary total hip arthroplasty - Minimal 5-year follow-up study

被引:23
作者
Han, CD [1 ]
Choe, WS [1 ]
Yoo, JH [1 ]
机构
[1] Yonsei Univ, Coll Med, Dept Orthopaed Surg, Seoul 120752, South Korea
关键词
cementless total hip; arthroplasty; polyethylene wear; particulate wear debris; osteolysis; loosening;
D O I
10.1016/S0883-5403(99)90227-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We measured polyethylene wear radiographically in 85 cementless primary total hip arthroplasties. All patients were followed up for at least 5 years (mean, 6.3 years). The mean extent of linear wear was 0.88 mm, and the mean rate of linear wear was 0.14 mm/y. The mean volumetric wear was 539.42 mm(3), and the mean rate of volumetric wear was 87.69 mm(3)/y. Polyethylene wear was correlated inversely with the patients' age only and was not influenced by other clinical factors. Osteolysis developed in 34 femurs and in ii acetabula. Radiographic loosening was found in 2 cases in the femur and none in the acetabulum. Polyethylene wear was significantly correlated with frequency and size of osteolytic lesions. Therefore, polyethylene wear was considered one of the major causes of osteolysis, but it was not directly associated with prosthesis loosening. To reduce the extent of polyethylene wear, the material characteristics of polyethylene should be improved, or more durable articulating materials should he developed.
引用
收藏
页码:714 / 723
页数:10
相关论文
共 32 条
[1]  
[Anonymous], CLIN ORTHOPAEDICS
[2]   Periprosthetic bone resorption - Particles versus movement [J].
Aspenberg, P ;
Herbertsson, P .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (04) :641-646
[3]  
BANKSTON AB, 1995, CLIN ORTHOP RELAT R, P37
[4]   POLYETHYLENE WEAR IN CEMENTED METAL-BACKED ACETABULAR CUPS [J].
CATES, HE ;
FARIS, PM ;
KEATING, EM ;
RITTER, MA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02) :249-253
[5]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[6]   Measurement of polyethylene wear in acetabular components inserted with and without cement - A randomized trial [J].
Devane, PA ;
Robinson, EJ ;
Bourne, RB ;
Rorabeck, CH ;
Nayak, NN ;
Horne, JG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (05) :682-689
[7]  
Devane PA, 1995, CLIN ORTHOP RELAT R, V319, P303
[8]   A QUANTITATIVE-EVALUATION OF PERIPROSTHETIC BONE-REMODELING AFTER CEMENTLESS TOTAL HIP-ARTHROPLASTY [J].
ENGH, CA ;
MCGOVERN, TF ;
BOBYN, JD ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) :1009-1020
[9]   POROUS-COATED HIP-REPLACEMENT - THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL-RESULTS [J].
ENGH, CA ;
BOBYN, JD ;
GLASSMAN, AH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01) :45-55
[10]  
GomezBarrena E, 1996, AAOS INSTR COURS LEC, V45, P187