Serial intravascular ultrasound predictors of restenosis at the margins of Palmaz-Schatz stents

被引:45
作者
Hoffmann, R [1 ]
Mintz, GS [1 ]
Kent, KM [1 ]
Satler, LF [1 ]
Pichard, AD [1 ]
Popma, JJ [1 ]
Leon, MB [1 ]
机构
[1] WASHINGTON HOSP CTR,INTRAVASC ULTRASOUND IMAGING & CARDIAC CATHETERIZ,WASHINGTON,DC 20010
关键词
D O I
10.1016/S0002-9149(97)00016-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To evaluate predictors of restenosis at margins of Palmaz-Schatz stents, intravascular ultrasound studies were performed after intervention and at follow-up (5.4 months) in 161 stented lesions. Of 301 stent margins, 77 (26%) were restenotic at follow-up (>50% late lumen loss). Intimal hyperplasia was greater for restenotic than for nonrestenotic stents margins. The dominant periprocedural predictor of stent margin restenosis was the plaque burden of the continuous reference segment.
引用
收藏
页码:951 / &
页数:4
相关论文
共 12 条
[1]  
Bailey Steven R., 1994, P893
[2]   A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE [J].
FISCHMAN, DL ;
LEON, MB ;
BAIM, DS ;
SCHATZ, RA ;
SAVAGE, MP ;
PENN, I ;
DETRE, K ;
VELTRI, L ;
RICCI, D ;
NOBUYOSHI, M ;
CLEMAN, M ;
HEUSER, R ;
ALMOND, D ;
TEIRSTEIN, PS ;
FISH, RD ;
COLOMBO, A ;
BRINKER, J ;
MOSES, J ;
SHAKNOVICH, A ;
HIRSHFELD, J ;
BAILEY, S ;
ELLIS, S ;
RAKE, R ;
GOLDBERG, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) :496-501
[3]   MECHANISMS OF RESTENOSIS AND REDILATION WITHIN CORONARY STENTS - QUANTITATIVE ANGIOGRAPHIC ASSESSMENT [J].
GORDON, PC ;
GIBSON, CM ;
COHEN, DJ ;
CARROZZA, JP ;
KUNTZ, RE ;
BAIM, DS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (05) :1166-1174
[4]  
HAUDE M, 1991, BRIT HEART J, V66, P337
[5]   Chronic arterial responses to stent implantation: A serial intravascular ultrasound analysis of Palmaz-Schatz stents in native coronary arteries [J].
Hoffmann, R ;
Mintz, GS ;
Popma, JJ ;
Satler, LF ;
Pichard, AD ;
Kent, KM ;
Walsh, C ;
Mackell, P ;
Leon, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 28 (05) :1134-1139
[6]   Patterns and mechanisms of in-stent restenosis - A serial intravascular ultrasound study [J].
Hoffmann, R ;
Mintz, GS ;
Dussaillant, GR ;
Popma, JJ ;
Pichard, AD ;
Satler, LF ;
Kent, KM ;
Griffin, J ;
Leon, MB .
CIRCULATION, 1996, 94 (06) :1247-1254
[7]   LUMINAL LOSS AND SITE OF RESTENOSIS AFTER PALMAZ-SCHATZ CORONARY STENT IMPLANTATION [J].
IKARI, Y ;
HARA, K ;
TAMURA, T ;
SAEKI, F ;
YAMAGUCHI, T .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (03) :117-120
[8]   INTRAVASCULAR ULTRASOUND ASSESSMENT OF THE MAGNITUDE AND MECHANISM OF CORONARY-ARTERY AND LUMEN TAPERING [J].
JAVIER, SP ;
MINTZ, GS ;
POPMA, JJ ;
PICHARD, AD ;
KENT, KM ;
SATLER, LF ;
LEON, MB .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 75 (02) :177-180
[9]   ATHEROSCLEROSIS IN ANGIOGRAPHICALLY NORMAL CORONARY-ARTERY REFERENCE SEGMENTS - AN INTRAVASCULAR ULTRASOUND STUDY WITH CLINICAL CORRELATIONS [J].
MINTZ, GS ;
PAINTER, JA ;
PICHARD, AD ;
KENT, KM ;
SATLER, LF ;
POPMA, JJ ;
CHUANG, YC ;
BUCHER, TA ;
SOKOLOWICZ, LE ;
LEON, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (07) :1479-1485
[10]   Arterial remodeling after coronary angioplasty - A serial intravascular ultrasound study [J].
Mintz, GS ;
Popma, JJ ;
Pichard, AD ;
Kent, KM ;
Satler, LF ;
Wong, SC ;
Hong, MK ;
Kovach, JA ;
Leon, MB .
CIRCULATION, 1996, 94 (01) :35-43