Prismatic imaging polarimeter calibration for the infrared spectral region

被引:24
作者
Kudenov, Michael W. [1 ]
Pezzaniti, Larry [2 ]
Dereniak, Eustace L. [1 ]
Gerhart, Grant R. [3 ]
机构
[1] Univ Arizona, Coll Opt Sci, Tucson, AZ 85721 USA
[2] Polaris Sensor Technol Inc, Huntsville, AL USA
[3] USA, TACOM, Warren, MI 48397 USA
关键词
Polarimeters - Infrared radiation;
D O I
10.1364/OE.16.013720
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The calibration of a complete Stokes birefringent prismatic imaging polarimeter (BPIP) in the MWIR is demonstrated. The BPIP technique, originally developed by K. Oka, is implemented with a set of four Yttrium Vanadate (YVO4) crystal prisms. A mathematical model for the polarimeter is presented in which diattenuation due to Fresnel effects and dichroism in the crystal are included. An improved polarimetric calibration technique is introduced to remove the diattenuation effects, along with the relative radiometric calibration required for the BPIP operating with a thermal background and large detector offsets. Data demonstrating emission polarization are presented from various blackbodies, which are compared to data from our Fourier transform infrared spectropolarimeter. (c) 2008 Optical Society of America.
引用
收藏
页码:13720 / 13737
页数:18
相关论文
共 12 条
[1]   LWIR polarimeter calibration [J].
Blumer, RV ;
Miller, MA ;
Howe, JD ;
Stevens, MA .
POLARIZATION ANALYSIS, MEASUREMENT, AND REMOTE SENSING IV, 2002, 4481 :37-45
[2]   Unpolarized calibration and nonuniformity correction for long-wave infrared microgrid imaging polarimeters [J].
Bowers, David L. ;
Boger, James K. ;
Wellems, L. David ;
Ortega, Steve E. ;
Fetrow, Matthew P. ;
Hubbs, John E. ;
Black, Wiley T. ;
Ratliff, Bradley M. ;
Tyo, J. Scott .
OPTICAL ENGINEERING, 2008, 47 (04)
[3]   Improved mid-infrared polarizers using yttrium vanadate [J].
DeShazer, LG .
POLARIZATION ANALYSIS, MEASUREMENT, AND REMOTE SENSING IV, 2002, 4481 :10-16
[4]   256x256 InSb focal plane arrays [J].
Gau, YT ;
Dai, LK ;
Yang, SP ;
Weng, PK ;
Huang, KS ;
Liu, YN ;
Chiang, CD ;
Jih, FW ;
Cherng, YT ;
Chang, H .
OPTOELECTRONIC MATERIALS AND DEVICES II, 2000, 4078 :467-479
[5]   Fourier transform channeled spectropolarimetry in the MWIR [J].
Kudenov, Michael W. ;
Hagen, Nathan A. ;
Dereniak, Eustace L. ;
Gerhart, Grant R. .
OPTICS EXPRESS, 2007, 15 (20) :12792-12805
[6]   A BLACK-BODY SOURCE OF RADIATION COVERING A WAVELENGTH RANGE FROM ULTRAVIOLET TO INFRARED [J].
LAPWORTH, KC ;
QUINN, TJ ;
ALLNUTT, LA .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1970, 3 (02) :116-+
[7]   High birefringence of the yttrium vanadate crystal in the middle wavelength infrared [J].
Luo, HT ;
Tkaczyk, T ;
Dereniak, EL ;
Oka, K ;
Sampson, R .
OPTICS LETTERS, 2006, 31 (05) :616-618
[8]   Compact complete imaging polarimeter using birefringent wedge prisms [J].
Oka, K ;
Kaneko, T .
OPTICS EXPRESS, 2003, 11 (13) :1510-1519
[9]  
PEZZANITI L, 2005, P SPIE, V2
[10]   Stabilization of a channeled spectropolarimeter by self-calibration [J].
Taniguchi, Atsushi ;
Oka, Kazuhiko ;
Okabe, Hiroshi ;
Hayakawa, Masayuki .
OPTICS LETTERS, 2006, 31 (22) :3279-3281