A low temperature ultrahigh vaccum scanning force microscope

被引:101
作者
Hug, HJ [1 ]
Stiefel, B [1 ]
van Schendel, PJA [1 ]
Moser, A [1 ]
Martin, S [1 ]
Güntherodt, HJ [1 ]
机构
[1] Univ Basel, Inst Phys, CH-4056 Basel, Switzerland
关键词
D O I
10.1063/1.1149970
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This article describes the design of a versatile ultrahigh vaccum (UHV) low temperature scanning force microscope system. The system allows scanning probe microscopy measurements at temperatures between 6 and 400 K and in magnetic fields up to 7 T. Cantilevers and samples can be prepared in UHV and transferred to the microscope. We describe some technical details of our system and present first measurements performed at different temperatures and in various scanning force microscopy operation modes. We demonstrate distortion free and calibrated images at temperatures ranging from 8 to 300 K, atomic resolution on NaCl at 7.6 K and various magnetic force microscopy images of vortices in high transition temperature superconductors. It is demonstrated that our instrumentation reaches the thermodynamically determined sensitivity limit. Using standard cantilevers force gradients in the 10(-6)N/m range, corresponding forces of about 10(-15)N can be measured. (C) 1999 American Institute of Physics. [S0034-6748(99)00909-0].
引用
收藏
页码:3625 / 3640
页数:16
相关论文
共 74 条
[1]  
AKIA J, 1995, REV SCI INSTRUM, V66, P2517
[2]   LOW-TEMPERATURE FORCE MICROSCOPE WITH ALL-FIBER INTERFEROMETER [J].
ALBRECHT, TR ;
GRUTTER, P ;
RUGAR, D ;
SMITH, DPE .
ULTRAMICROSCOPY, 1992, 42 :1638-1646
[3]   FREQUENCY-MODULATION DETECTION USING HIGH-Q CANTILEVERS FOR ENHANCED FORCE MICROSCOPE SENSITIVITY [J].
ALBRECHT, TR ;
GRUTTER, P ;
HORNE, D ;
RUGAR, D .
JOURNAL OF APPLIED PHYSICS, 1991, 69 (02) :668-673
[4]   Scanning force microscope with atomic resolution in ultrahigh vacuum and at low temperatures [J].
Allers, W ;
Schwarz, A ;
Schwarz, UD ;
Wiesendanger, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (01) :221-225
[5]  
BAMMERLIN M, 1995, THESIS U BASEL
[6]   Thermal analysis using a micromechanical calorimeter [J].
Berger, R ;
Gerber, C ;
Gimzewski, JK ;
Meyer, E ;
Guntherodt, HJ .
APPLIED PHYSICS LETTERS, 1996, 69 (01) :40-42
[7]   MAGNETIC DOMAIN-STRUCTURE IN ULTRATHIN FILMS [J].
BOCHI, G ;
HUG, HJ ;
PAUL, DI ;
STIEFEL, B ;
MOSER, A ;
PARASHIKOV, I ;
GUNTHERODT, HJ ;
OHANDLEY, RC .
PHYSICAL REVIEW LETTERS, 1995, 75 (09) :1839-1842
[8]   COMPUTER CORRECTION FOR DISTORTED STM IMAGES [J].
CAI, CZ ;
CHEN, XY ;
SHU, QQ ;
ZHENG, XL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (12) :5649-5652
[9]   VARIABLE-TEMPERATURE ULTRAHIGH-VACUUM ATOMIC-FORCE MICROSCOPE [J].
DAI, Q ;
VOLLMER, R ;
CARPICK, RW ;
OGLETREE, DF ;
SALMERON, M .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (11) :5266-5271
[10]   Growth and etching phenomena observed by STM/AFM on pulsed-laser deposited YBa2Cu3O7-delta films [J].
Dam, B ;
Koeman, NJ ;
Rector, JH ;
StaublePumpin, B ;
Poppe, U ;
Griessen, R .
PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 1996, 261 (1-2) :1-11