Comparing fixed points of the ITS-90 as realized in Canada, Argentina and Brazil

被引:4
作者
Ancsin, J [1 ]
Tischler, M
Petkovic, SG
Santiago, JFN
机构
[1] Natl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, Canada
[2] INTI, RA-1650 Buenos Aires, DF, Argentina
[3] INMETRO, Rio De Janeiro, Brazil
关键词
D O I
10.1088/0026-1394/38/6/11
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A set of fixed points maintained by national metrology institutes (NMIs) for calibrating platinum resistance thermometers (PRTs) in Canada (National Research Council, NRC), Argentina (Instituto Nacional de Tecnologia Industrial, INTI) and Brazil (Instituto Nacional de Metrologia, Normalizacao e Qualidade Industrial, INMETRO) were compared using two methods. A set of portable cells was used in one method, and in the other a calibrated PRT. The fixed-point differences were as follows. For the INTI the differences DeltaT = T(INTI) -T(NRC) found at the Ag, Al, Zn, Sn, In, Ga and Hg points were 3.5 mK, 2.8 mK, 1.8 mK, -0.5 mK, 0.2 mK, -0.2 mK and -0.5 mK, respectively. For the INMETRO the differences DeltaT = T(INMETRO) - T(NRC) at the above fixed points are 0.7 mK, 3.3 mK, -1.3 mK, -1.9 mK, -0.3 mK, -0.1 mK, -0.4 mK, respectively, and 0.8 mK at the Ar point.
引用
收藏
页码:555 / 566
页数:12
相关论文
共 12 条
[1]   Concerning the consistency of fixed points of the temperature scale [J].
Ancsin, J .
METROLOGIA, 1996, 32 (04) :295-300
[2]   INTERCOMPARING THE FREEZING BEHAVIOR OF SILVER CELLS OF THE NRC AND THE ITRI [J].
ANCSIN, J ;
YEH, TI ;
LIN, J ;
CHU, YH .
METROLOGIA, 1993, 30 (01) :23-30
[3]   THE EFFECTS OF ANNEALING ON THE RESISTANCE OF HIGH-TEMPERATURE PRTS AT THE TRIPLE POINT OF WATER AFTER HIGH-TEMPERATURE USE [J].
ANCSIN, J ;
ZHANG, Y .
METROLOGIA, 1994, 31 (02) :85-91
[4]   Comparison of realizations of the thermometric fixed points of the ITS-90 in the range -38 degrees C to 420 degrees C between Canada and Mexico [J].
Ancsin, J ;
MendezLango, E .
METROLOGIA, 1996, 33 (04) :415-420
[5]   Equilibrium melting curves of relatively pure and doped silver samples [J].
Ancsin, J .
METROLOGIA, 2001, 38 (03) :229-235
[6]   The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using platinum resistance thermometers [J].
Ancsin, J ;
Mendez-Lango, E .
METROLOGIA, 1999, 36 (02) :117-139
[7]   Equilibrium melting curves of silver using high-temperature calorimeters [J].
Ancsin, J .
METROLOGIA, 2001, 38 (01) :1-7
[8]  
ANCSIN J, 1999, P TEMP THERM MEAS IN, P253
[9]   Comparisons of some NIST fixed-point cells with similar cells of other standards laboratories [J].
Mangum, BW ;
Pfeiffer, ER ;
Strouse, GF ;
ValenciaRodriguez, J ;
Lin, JH ;
Yeh, TI ;
Marcarino, P ;
Dematteis, R ;
Liu, Y ;
Zhao, Q ;
Ince, AT ;
Cakiroglu, F ;
Nubbemeyer, HG ;
Jung, HJ .
METROLOGIA, 1996, 33 (03) :215-225
[10]   On the International Temperature Scale of 1990 (ITS-90) - Part II: Recommended techniques for comparisons, at the highest level of accuracy, of fixed-point cells used for contact thermometry [J].
Mangum, BW ;
Bloembergen, P ;
Chattle, MV ;
Fellmuth, B ;
Marcarino, P ;
Pokhodun, I .
METROLOGIA, 1999, 36 (02) :79-88