All-solid-state planar miniature ion-selective chloride electrode

被引:103
作者
Zielinska, R
Mulik, E
Michalska, A
Achmatowicz, S
Maj-Zurawska, M
机构
[1] Univ Warsaw, Dept Chem, PL-00093 Warsaw, Poland
[2] Inst Technol Elect Mat, PL-01919 Warsaw, Poland
关键词
all-solid-state potentiometric sensor; screen-printed electrodes; chloride sensor; poly(pyrrole);
D O I
10.1016/S0003-2670(01)01407-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
All-solid-state ion-selective electrodes with plastic membrane (poly(vinyl chloride) (PVC), bis(2-ethylhexyl) sebacate (DOS), methyltri-n-tetradecylammonium chloride (MTTACl)), a conducting poly(pyrrole) (PPy) film doped either with chloride ions (PPyCl) or hexacyanoferrate(II) ions (PPyFeCN), and glassy carbon (GC) or screen-printed graphite layer (S-PG) as an inner electric contact were investigated. All the electrodes show close to Nernstian response, but their lifetimes vary. The at least 2-months lifetime of screen-printed electrodes is only achieved for the electrodes containing PPyFeCN (cation-exchanging film). Shorter lifetime of other screen-printed electrodes, i.e. without PPy, or with PPyCl (anion-exchanging film), was attributed to the diffusion of anionic products of the hydrolysis of organic components of the graphite paste used to prepare the electric contact. The properties of miniature, screen-printed electrodes comprising PPyFeCN solid contact, were comparable to those ion-selective electrodes with PPy solid contact (regardless the ion-exchanging characteristic of the polymer) deposited on GC electric contact. (C) 2002 Elsevier Science B.V All rights reserved.
引用
收藏
页码:243 / 249
页数:7
相关论文
共 18 条
[1]   Potential stability of all-solid-state ion-selective electrodes using conducting polymers as ion-to-electron transducers [J].
Bobacka, J .
ANALYTICAL CHEMISTRY, 1999, 71 (21) :4932-4937
[2]   SINGLE PIECE ALL-SOLID-STATE ION-SELECTIVE ELECTRODE [J].
BOBACKA, J ;
LINDFORS, T ;
MCCARRICK, M ;
IVASKA, A ;
LEWENSTAM, A .
ANALYTICAL CHEMISTRY, 1995, 67 (20) :3819-3823
[3]   ALL-SOLID-STATE POLY(VINYL CHLORIDE) MEMBRANE ION-SELECTIVE ELECTRODES WITH POLY(3-OCTYLTHIOPHENE) SOLID INTERNAL CONTACT [J].
BOBACKA, J ;
MCCARRICK, M ;
LEWENSTAM, A ;
IVASKA, A .
ANALYST, 1994, 119 (09) :1985-1991
[4]   ALL-SOLID-STATE SODIUM-SELECTIVE ELECTRODE BASED ON A CALIXARENE IONOPHORE IN A POLY(VINYL CHLORIDE) MEMBRANE WITH A POLYPYRROLE SOLID CONTACT [J].
CADOGAN, A ;
GAO, ZQ ;
LEWENSTAM, A ;
IVASKA, A ;
DIAMOND, D .
ANALYTICAL CHEMISTRY, 1992, 64 (21) :2496-2501
[5]   MICROFABRICATED SENSOR ARRAYS SENSITIVE TO PH AND K+ FOR IONIC DISTRIBUTION MEASUREMENTS IN THE BEATING HEART [J].
COSOFRET, VV ;
ERDOSY, M ;
JOHNSON, TA ;
BUCK, RP ;
ASH, RB ;
NEUMAN, MR .
ANALYTICAL CHEMISTRY, 1995, 67 (10) :1647-1653
[6]   ALL-SOLID-STATE CHLORIDE-SELECTIVE ELECTRODE WITH POLY(PYRROLE) SOLID CONTACT [J].
HULANICKI, A ;
MICHALSKA, A .
ELECTROANALYSIS, 1995, 7 (07) :692-693
[7]   MICROFIBER MATRIX-SUPPORTED ION-SELECTIVE PVC MEMBRANES [J].
KNOLL, M ;
CAMMANN, K ;
DUMSCHAT, C ;
BORCHARDT, M ;
HOGG, G .
SENSORS AND ACTUATORS B-CHEMICAL, 1994, 20 (01) :1-5
[8]   Disposable strip potentiometric electrodes with solvent-polymeric ion-selective membranes fabricated using screen-printing technology [J].
Koncki, R ;
Glab, S ;
Dziwulska, J ;
Palchetti, I ;
Mascini, M .
ANALYTICA CHIMICA ACTA, 1999, 385 (1-3) :451-459
[9]   COATED FILM ELECTRODES [J].
LEMKE, U ;
CAMMANN, K .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1989, 335 (07) :852-854
[10]  
MEIER PC, 1980, MED BIOL APPLICATION, pCH2