Rapid prototyping (RP) is a relatively new technology that produces physical models by selectively solidifying UV-sensitive liquid resin using a laser beam. The technology has gained a great amount of attentin, particularly in oral and maxillofacial surgery. An important issue in RP applications in this field is how to obtain RP models of the required accuracy. We investigated errors generated during the production of medical RP models, and identified the factors that caused dimensional errors in each production phase. The errors were mainly due to the volume-averaging effect, threshold value, and difficulty in the exact replication of landmark locations. We made 16 linear measurements on a dry skull, a replicated three-dimensional (3-D) visual (STL) model, and an RP model. The results showed that the absolute mean deviation between the original dry skull and the RP model over the 16 linear measurements was 0.62 +/- 0.35 mm (0.56 +/- 0.39%), which is smaller than values reported in previous studies. A major emphasis is placed on the dumb-bell effect. Classifying measurements as internal and external measurements, we observed that the effect of an inadequate threshold value differs with the type of measurement.
引用
收藏
页码:23 / 32
页数:10
相关论文
共 14 条
[1]  
ANDREWS JC, 1994, AM J OTOL, V15, P335
[2]  
Barker T M, 1994, Australas Radiol, V38, P106, DOI 10.1111/j.1440-1673.1994.tb00146.x
[3]   STEREOLITHOGRAPHY IN ORAL AND MAXILLOFACIAL OPERATION PLANNING [J].
BILL, JS ;
REUTHER, JF ;
DITTMANN, W ;
KUBLER, N ;
MEIER, JL ;
PISTNER, H ;
WITTENBERG, G .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1995, 24 (01) :98-103
[4]   Custom-made cast titanium implants produced with CAD/CAM for the reconstruction of cranium defects [J].
Heissler, E ;
Fischer, FS ;
Bolouri, S ;
Lehmann, T ;
Mathar, W ;
Gebhardt, A ;
Lanksch, W ;
Bier, J .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1998, 27 (05) :334-338
[5]   PEDIATRIC CRANIOFACIAL SURGERY - COMPARISON OF MILLING AND STEREOLITHOGRAPHY FOR 3D-MODEL MANUFACTURING [J].
KLEIN, HM ;
SCHNEIDER, W ;
ALZEN, G ;
VOY, ED ;
GUNTHER, RW .
PEDIATRIC RADIOLOGY, 1992, 22 (06) :458-460
[6]   A comparison of three-dimensional computed tomography scans and stereolithographic models for evaluation of craniofacial anomalies [J].
Kragskov, J ;
SindetPedersen, S ;
Gyldensted, C ;
Jensen, KL .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1996, 54 (04) :402-411
[7]  
LAMBRECHT JT, 1995, MODELING TECHNOLOGY, P61
[8]   REPRODUCIBILITY OF 3-DIMENSIONAL CT-ASSISTED MODEL PRODUCTION IN THE MAXILLOFACIAL AREA [J].
LILL, W ;
SOLAR, P ;
ULM, C ;
WATZEK, G ;
BLAHOUT, R ;
MATEJKA, M .
BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 1992, 30 (04) :233-236
[9]   3D virtual endoscopy of the upper airway: Optimization of the scan parameters in a cadaver phantom and clinical assessment [J].
Rodenwaldt, J ;
Kopka, L ;
Roedel, R ;
Margas, A ;
Grabbe, E .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1997, 21 (03) :405-411
[10]   The value of stereolithographic models for preoperative diagnosis of craniofacial deformities and planning of surgical corrections [J].
Sailer, HF ;
Haers, PE ;
Zollikofer, CPE ;
Warnke, T ;
Carls, FR ;
Stucki, P .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1998, 27 (05) :327-333