Hybrid ordered phase encoding (HOPE): An improved approach for respiratory artifact reduction

被引:52
作者
Jhooti, P [1 ]
Wiesmann, F [1 ]
Taylor, AM [1 ]
Gatehouse, PD [1 ]
Yang, GZ [1 ]
Keegan, J [1 ]
Pennell, DJ [1 ]
Firmin, DN [1 ]
机构
[1] Royal Brompton & Natl Heart Hosp, Magnet Resonance Unit, London SW3 6NP, England
关键词
motion artifacts; phase-encode reordering; respiratory monitoring; cardiac imaging;
D O I
10.1002/jmri.1880080428
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Respiration causes continuous change in cardiac position, which leads to image degradation. Phase-encode reordering methods are often used to reduce these artifacts. An improved method for suppressing motion artifacts by reordering the acquisition of k space has been developed that is less sensitive to change of breathing patterns and bulk movement. We describe the theory behind the new approach and compare its results with those of existing methods by use of a phantom with simulated and actual acquired breathing patterns. The comparison was also made in vivo: cardiac scans were performed in 15 subjects with image planes that are known to be particularly susceptible to respiratory artifact. A significant improvement in image quality was achieved compared with conventional nonreordered and existing reordering methods.
引用
收藏
页码:968 / 980
页数:13
相关论文
共 17 条
  • [1] RESPIRATORY ORDERED PHASE ENCODING (ROPE) - A METHOD FOR REDUCING RESPIRATORY MOTION ARTIFACTS IN MR IMAGING
    BAILES, DR
    GILDERDALE, DJ
    BYDDER, GM
    COLLINS, AG
    FIRMIN, DN
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (04) : 835 - 838
  • [2] THE EFFECT OF RESPIRATION ON THE CONTRAST AND SHARPNESS OF LIVER-LESIONS IN MRI
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (01) : 1 - 7
  • [3] SEGMENTED TURBOFLASH - METHOD FOR BREATH-HOLD MR IMAGING OF THE LIVER WITH FLEXIBLE CONTRAST
    EDELMAN, RR
    WALLNER, B
    SINGER, A
    ATKINSON, DJ
    SAINI, S
    [J]. RADIOLOGY, 1990, 177 (02) : 515 - 521
  • [4] MAGNETIC-RESONANCE IMAGING WITH RESPIRATORY GATING - TECHNIQUES AND ADVANTAGES
    EHMAN, RL
    MCNAMARA, MT
    PALLACK, M
    HRICAK, H
    HIGGINS, CB
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 143 (06) : 1175 - 1182
  • [5] DIAPHRAGMATIC MOTION - FAST GRADIENT-RECALLED-ECHO MR-IMAGING IN HEALTHY-SUBJECTS
    GIERADA, DS
    CURTIN, JJ
    ERICKSON, SJ
    PROST, RW
    STRANDT, JA
    GOODMAN, LR
    [J]. RADIOLOGY, 1995, 194 (03) : 879 - 884
  • [6] GLOVER GH, 1988, Patent No. 4663691
  • [7] HAACKE E M, 1986, Magnetic Resonance Imaging, V4, P359, DOI 10.1016/0730-725X(86)91046-5
  • [8] SIGNAL-TO-NOISE, RESOLUTION, AND BIAS FUNCTION-ANALYSIS OF ASYMMETRIC SAMPLING WITH ZERO-PADDED MAGNITUDE FT RECONSTRUCTION
    HURST, GC
    HUA, JM
    SIMONETTI, OP
    DUERK, JL
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 27 (02) : 247 - 269
  • [9] ALTERED PHASE-ENCODING ORDER FOR REDUCED SENSITIVITY TO MOTION IN 3-DIMENSIONAL MR IMAGING
    KORIN, HW
    RIEDERER, SJ
    BAMPTON, AEH
    EHMAN, RL
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (06): : 687 - 693
  • [10] NMR FOURIER ZEUGMATOGRAPHY
    KUMAR, A
    WELTI, D
    ERNST, RR
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1975, 18 (01) : 69 - 83