Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas MD Anderson Cancer Center, Proton Therapy Center, Houston

被引:237
作者
Gillin, Michael T. [1 ]
Sahoo, Narayan [1 ]
Bues, Martin [1 ]
Ciangaru, George [1 ]
Sawakuchi, Gabriel [1 ]
Poenisch, Falk [1 ]
Arjomandy, Bijan [1 ]
Martin, Craig [1 ]
Titt, Uwe [1 ]
Suzuki, Kazumichi [1 ]
Smith, Alfred R. [1 ]
Zhu, X. Ronald [1 ]
机构
[1] UT MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
spot scanning proton beam; clinical commissioning;
D O I
10.1118/1.3259742
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: To describe a summary of the clinical commissioning of the discrete spot scanning proton beam at the Proton Therapy Center, Houston (PTC-H). Methods: Discrete spot scanning system is composed of a delivery system (Hitachi ProBeat), an electronic medical record (Mosaiq V 1.5), and a treatment planning system (TPS) (Eclipse V 8.1). Discrete proton pencil beams (spots) are used to deposit dose spot by spot and layer by layer for the proton distal ranges spanning from 4.0 to 30.6 g/cm(2) and over a maximum scan area at the iso-center of 30 X 30 cm(2). An arbitrarily chosen reference calibration condition has been selected to define the monitor units (MUs). Using radiochromic film and ion chambers, the authors have measured spot positions, the spot sizes in air, depth dose curves, and profiles for proton beams with various energies in water, and studied the linearity of the dose monitors. In addition to dosimetric measurements and TPS modeling, significant efforts were spent in testing information flow and recovery of the delivery system from treatment interruptions. Results: The main dose monitors have been adjusted such that a specific amount of charge is collected in the monitor chamber corresponding to a single MU, following the IAEA TRS 398 protocol under a specific reference condition. The dose monitor calibration method is based on the absolute dose per MU, which is equivalent to the absolute dose per particle, the approach used by other scanning beam institutions. The full width at half maximum for the spot size in air varies from approximately 1.2 cm for 221.8 MeV to 3.4 cm for 72.5 MeV. The measured versus requested 90% depth dose in water agrees to within 1 mm over ranges of 4.0-30.6 cm. The beam delivery interlocks perform as expected, guarantying the safe and accurate delivery of the planned dose. Conclusions: The dosimetric parameters of the discrete spot scanning proton beam have been measured as part of the clinical commissioning program, and the machine is found to function in a safe manner, making it suitable for patient treatment. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3259742]
引用
收藏
页码:154 / 163
页数:10
相关论文
共 10 条
[1]
Andreo P, 2000, IAEA TECHNICAL REPOR
[2]
Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking [J].
Farr, J. B. ;
Mascia, A. E. ;
Hsi, W. -C. ;
Allgower, C. E. ;
Jesseph, F. ;
Schreuder, A. N. ;
Wolanski, M. ;
Nichiporov, D. F. ;
Anferov, V. .
MEDICAL PHYSICS, 2008, 35 (11) :4945-4954
[3]
RANGE MODULATORS FOR PROTONS AND HEAVY-IONS [J].
KOEHLER, AM ;
SCHNEIDER, RJ ;
SISTERSON, JM .
NUCLEAR INSTRUMENTS & METHODS, 1975, 131 (03) :437-440
[4]
Intensity modulated proton therapy: A clinical example [J].
Lomax, AJ ;
Boehringer, T ;
Coray, A ;
Egger, E ;
Goitein, G ;
Grossmann, M ;
Juelke, P ;
Lin, S ;
Pedroni, E ;
Rohrer, B ;
Roser, W ;
Rossi, B ;
Siegenthaler, B ;
Stadelmann, O ;
Stauble, H ;
Vetter, C ;
Wisser, L .
MEDICAL PHYSICS, 2001, 28 (03) :317-324
[5]
Reference dosimetry in a scanned pulsed proton beam using ionisation chambers and a Faraday cup [J].
Lorin, Stefan ;
Grusell, Erik ;
Tilly, Nina ;
Medin, Joakim ;
Kimstrand, Peter ;
Glimelius, Bengt .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (13) :3519-3529
[6]
Paganetti H, 2005, MED RAD DIA IMG, P345
[7]
Experimental characterization and physical modelling of the dose distribution of scanned proton pencil beams [J].
Pedroni, E ;
Scheib, S ;
Böhringer, T ;
Coray, A ;
Grossmann, M ;
Lin, S ;
Lomax, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (03) :541-561
[8]
THE 200-MEV PROTON THERAPY PROJECT AT THE PAUL-SCHERRER-INSTITUTE - CONCEPTUAL DESIGN AND PRACTICAL REALIZATION [J].
PEDRONI, E ;
BACHER, R ;
BLATTMANN, H ;
BOHRINGER, T ;
CORAY, A ;
LOMAX, A ;
LIN, SX ;
MUNKEL, G ;
SCHEIB, S ;
SCHNEIDER, U ;
TOUROVSKY, A .
MEDICAL PHYSICS, 1995, 22 (01) :37-53
[9]
The M. D. Anderson proton therapy system [J].
Smith, Alfred ;
Gillin, Michael ;
Bues, Martin ;
Zhu, X. Ronald ;
Suzuki, Kazumichi ;
Mohan, Radhe ;
Woo, Shiao ;
Lee, Andrew ;
Komaki, Ritsko ;
Cox, James ;
Hiramoto, Kazuo ;
Akiyama, Hiroshi ;
Ishida, Takayuki ;
Sasaki, Toshie ;
Matsuda, Koji .
MEDICAL PHYSICS, 2009, 36 (09) :4068-4083
[10]
Development and verification of the pulsed scanned proton beam at The Svedberg Laboratory in Uppsala [J].
Tilly, Nina ;
Grusell, Erik ;
Kimstrand, Peter ;
Lorin, Stefan ;
Gajewski, Konrad ;
Pettersson, Mikael ;
Backlund, Andreas ;
Glimelius, Bengt .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (10) :2741-2754