Comparison of two digital radio stereometric analysis methods in the determination of femoral head penetration in a total hip replacement phantom

被引:36
作者
Bragdon, CR
Estok, DM
Malchau, H
Kärrholm, J
Yuan, XH
Bourne, R
Veldhoven, J
Harris, WH
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med,Adult Reconstruct Unit, Dept Orthopaed Surg,Orthopaed Biomech & Biomat La, Boston, MA 02114 USA
[2] Univ Western Ontario, London, ON N6K 1C7, Canada
[3] Univ Gothenburg, Inst Surg Sci, Dept Orthopaed, Gothenburg, Sweden
关键词
radiostereometric analysis; polyethylene wear; total hip arthroplasty; accuracy;
D O I
10.1016/j.orthres.2003.09.004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Radiostereometric analysis (RSA) has been used extensively to evaluate the magnitude and direction of penetration of the femoral head into the acetabular component of a total hip replacement as a result of polyethylene wear and creep. The accuracy and precision of an RSA study depends on several factors, including the radiographic technique, the analytical software, and the positioning of the tantalum markers. This study had three sequential purposes. First, an in vitro phantom model was used to quantify the accuracy and precision of digital images versus conventional radiography in RSA measurements of penetration of the femoral head into the acetabular shell in a total hip replacement. The Umea RSA software package was used for analysis of both the conventional films, which were digitized at a resolution of 300 DPI, and digital radiographs, which were converted from a DICOM format at a resolution of 218 DPI. Digital radiography was found to be superior. Next, two methods of RSA analysis currently in use for determining femoral head penetration into polyethylene of total joint replacements were compared. Using the phantom model, we compared the Umea RSA system (Biomedical Innovations AB) to the RSA-CMS (RSA Clinical Measurement Solution) and in both cases used the digital radiographs. The Umea RSA system was found to be superior. Finally, two methods of marking the position of the acetabular component with tantalum beads were compared: one in which beads were inserted into previously described towers protruding from the back of the acetabular shell and another in which beads were placed into the peripheral flange of the polyethylene liner using the Umea RSA analysis system of the digital radiographs. The results using the two marker configurations were similar. (C) 2003 Orthopaedic Research Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:659 / 664
页数:6
相关论文
共 14 条
[1]
Cement migration after THR -: A comparison of Charnley Elite and Exeter femoral stems using RSA [J].
Alfaro-Adrián, J ;
Gill, HS ;
Murray, DW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01) :130-134
[2]
Altman DG., 2013, STAT CONFIDENCE CONF
[3]
INSTABILITY AND WEAR OF TOTAL HIP PROSTHESES DETERMINED WITH ROENTGEN STEREOPHOTOGRAMMETRY [J].
BALDURSSON, H ;
EGUND, N ;
HANSSON, LI ;
SELVIK, G .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1979, 95 (04) :257-263
[4]
The precision of radiostereometric measurements.: Manual vs. digital measurements [J].
Börlin, N ;
Thien, T ;
Kärrholm, J .
JOURNAL OF BIOMECHANICS, 2002, 35 (01) :69-79
[5]
BORLIN N, 2000, THESIS UMEA U SWEDEN
[6]
Experimental assessment of precision and accuracy of radio stereometric analysis for the determination of polyethylene wear in a total hip replacement model [J].
Bragdon, CR ;
Malchau, H ;
Yuan, XH ;
Perinchief, R ;
Kärrholm, J ;
Börlin, N ;
Estok, DM ;
Harris, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (04) :688-695
[7]
Karrholm J, 1997, CLIN ORTHOP RELAT R, P94
[8]
Migration and wear of hydroxyapatite-coated press-fit cups in revision hip arthroplasty -: A radiostereometric study [J].
Nivbrant, B ;
Kärrholm, J .
JOURNAL OF ARTHROPLASTY, 1997, 12 (08) :904-912
[9]
Wear in uncemented porous and cemented polyethylene sockets -: A randomised, radiostereometric study [J].
Önsten, I ;
Carlsson, ÅS ;
Besjakov, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (02) :345-350
[10]
Ranstam J, 2000, ACTA ORTHOP SCAND, V71, P106