Prospective and Retrospective High Order Eddy Current Mitigation for Diffusion Weighted Echo Planar Imaging

被引:27
作者
Xu, Dan [1 ]
Maier, Joseph K. [2 ]
King, Kevin F. [1 ]
Collick, Bruce D. [2 ]
Wu, Gaohong [2 ]
Peters, Robert D. [1 ]
Hinks, R. Scott [1 ]
机构
[1] Gen Elect Healthcare, Appl Sci Lab, Milwaukee, WI USA
[2] Gen Elect Healthcare, MR Engn, Milwaukee, WI USA
关键词
diffusion MRI; EPI; artifacts; distortion; eddy currents; CURRENT-INDUCED ARTIFACTS; SPIN-ECHO; TREATMENT RESPONSE; IMAGES; DISTORTIONS; MRI; GRADIENTS; MOTION; COMPENSATION; TRACTOGRAPHY;
D O I
10.1002/mrm.24589
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
PurposeThe proposed method is aimed at reducing eddy current (EC) induced distortion in diffusion weighted echo planar imaging, without the need to perform further image coregistration between diffusion weighted and T2 images. These ECs typically have significant high order spatial components that cannot be compensated by preemphasis. Theory and MethodsHigh order ECs are first calibrated at the system level in a protocol independent fashion. The resulting amplitudes and time constants of high order ECs can then be used to calculate imaging protocol specific corrections. A combined prospective and retrospective approach is proposed to apply correction during data acquisition and image reconstruction. ResultsVarious phantom, brain, body, and whole body diffusion weighted images with and without the proposed method are acquired. Significantly reduced image distortion and misregistration are consistently seen in images with the proposed method compared with images without. ConclusionThe proposed method is a powerful (e.g., effective at 48 cm field of view and 30 cm slice coverage) and flexible (e.g., compatible with other image enhancements and arbitrary scan plane) technique to correct high order ECs induced distortion and misregistration for various diffusion weighted echo planar imaging applications, without the need for further image post processing, protocol dependent prescan, or sacrifice in signal-to-noise ratio. Magn Reson Med 70:1293-1305, 2013. (c) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:1293 / 1305
页数:13
相关论文
共 67 条
[1]
Elimination of eddy current artifacts in diffusion-weighted echo-planar images: The use of bipolar gradients [J].
Alexander, AL ;
Tsuruda, JS ;
Parker, DL .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (06) :1016-1021
[2]
Andersson J., 2012, P 20 ANN M ISMRM MEL, P2426
[3]
A model-based method for retrospective correction of geometric distortions in diffusion-weighted EPI [J].
Andersson, JLR ;
Skare, S .
NEUROIMAGE, 2002, 16 (01) :177-199
[4]
How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging [J].
Andersson, JLR ;
Skare, S ;
Ashburner, J .
NEUROIMAGE, 2003, 20 (02) :870-888
[5]
[Anonymous], 2015, Electromagnetic Theory
[6]
Diffusion Magnetic Resonance Imaging of the Breast [J].
Arantes Pereira, Fernanda Philadelpho ;
Martins, Gabriela ;
Carvalhaes de Oliveira, Raquel de Vasconcellos .
MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2011, 19 (01) :95-+
[7]
MR DIFFUSION TENSOR SPECTROSCOPY AND IMAGING [J].
BASSER, PJ ;
MATTIELLO, J ;
LEBIHAN, D .
BIOPHYSICAL JOURNAL, 1994, 66 (01) :259-267
[10]
On the use of water phantom images to calibrate and correct eddy current induced artefacts in MR diffusion tensor imaging [J].
Bastin, ME ;
Armitage, PA .
MAGNETIC RESONANCE IMAGING, 2000, 18 (06) :681-687