MRI of the brain is safe in patients implanted with the vagus nerve stimulator

被引:27
作者
Benbadis, SR
Nyhenhuis, J
Tatum, WO
Murtagh, FR
Gieron, M
Vale, FL
机构
[1] Univ S Florida, Coll Med, Dept Neurol, Comprehens Epilepsy Program, Tampa, FL 33606 USA
[2] Univ S Florida, Coll Med, Dept Neurol Surg, Tampa, FL 33606 USA
[3] Tampa Gen Hosp, Comprehens Epilepsy Program, Tampa, FL 33606 USA
[4] Purdue Univ, Sch Elect & Comp Engn, W Lafayette, IN 47907 USA
[5] Univ S Florida, Coll Med, Dept Radiol, Tampa, FL USA
[6] Univ S Florida, Coll Med, Dept Paediat, Tampa, FL USA
来源
SEIZURE-EUROPEAN JOURNAL OF EPILEPSY | 2001年 / 10卷 / 07期
关键词
vagus nerve stimulation; MRI;
D O I
10.1053/seiz.2001.0540
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Metallic devices generally represent a contra-indication for MRI scanning. Based on laboratory testing, the neuro cybernetic prosthesis (NCP) is labelled MRI compatible when used with a send and receive head coil. However, there are no published clinical data to support the safety of brain MRI in patients with the NCP. Our objective was to report clinical experience with such a population. We questioned 40 centres that had implanted the NCP system as of 10/1/99. If MRI had been performed on any vagus nerve stimulator patients, we collected information on these patients, the MRI technique used, any events noted during the scan, including both subjective reports (by the patient), and observable (objective) changes noted by the staff. Twelve centres (30%) responded. Over a time period of 3 years, there were a total of 27 MRI scans performed in 25 patients. All scanners were 1.5 T. A head coil was used in 26 scans, and a body coil in one. The indications for the scans were diverse. Seven were related to the epilepsy, including aetiology or pre-surgical evaluation. Others were unrelated, including brain tumours, cerebral haematoma, vasculitis, headaches, and head trauma. Three scans were performed with the stimulator on, while 24 were performed with the stimulator off. One patient had a mild objective voice change for several minutes. No other objective changes were noted in any of the patients. One 11-year old reported chest pain while experiencing severe claustrophobia. Twenty-five patients denied any discomfort around the lead or the generator. We conclude that this clinical series supports the safety of routine brain MRI using a send and receive head coil in patients implanted with the NCP System. (C) 2001 BEA Trading Ltd.
引用
收藏
页码:512 / 515
页数:4
相关论文
共 11 条
[1]   Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes [J].
Achenbach, S ;
Moshage, W ;
Diem, B ;
Bieberle, T ;
Schibgilla, V ;
Bachmann, K .
AMERICAN HEART JOURNAL, 1997, 134 (03) :467-473
[2]   When drugs don't work - An algorithmic approach to medically intractable epilepsy [J].
Benbadis, SR ;
Tatum, WO ;
Vale, FL .
NEUROLOGY, 2000, 55 (12) :1780-1784
[3]   RF heating of implanted spinal fusion stimulator during magnetic resonance imaging [J].
Chou, CK ;
McDougall, JA ;
Chan, KW .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (05) :367-373
[4]   A RANDOMIZED CONTROLLED TRIAL OF CHRONIC VAGUS NERVE-STIMULATION FOR TREATMENT OF MEDICALLY INTRACTABLE SEIZURES [J].
GEORGE, R ;
SONNEN, A ;
UPTON, A ;
SALINSKY, M ;
RISTANOVIC, R ;
BERGEN, D ;
MIRZA, W ;
ROSENFELD, W ;
NARITOKU, D ;
MANONESPAILLAT, R ;
BAROLAT, G ;
WILLIS, J ;
STEFAN, H ;
TREIG, T ;
HUFNAGEL, A ;
KUZNIECKY, R ;
UTHMAN, B ;
WILDER, BJ ;
AUGUSTINSSON, L ;
BENMENACHEM, E ;
RAMSAY, E ;
WERNICKE, JF ;
TARVER, WB .
NEUROLOGY, 1995, 45 (02) :224-230
[5]   Vagus nerve stimulation therapy for partial-onset seizures - A randomized active-control trial [J].
Handforth, A ;
DeGiorgio, CM ;
Schachter, SC ;
Uthman, BM ;
Naritoku, DK ;
Tecoma, ES ;
Henry, TR ;
Collins, SD ;
Vaughn, BV ;
Gilmartin, RC ;
Labar, DR ;
Morris, GL ;
Salinsky, MC ;
Osorio, I ;
Ristanovic, RK ;
Labiner, DM ;
Jones, JC ;
Murphy, JV ;
Ney, GC ;
Wheless, JW .
NEUROLOGY, 1998, 51 (01) :48-55
[6]  
Nyhenhuis J, 1997, EPILEPSIA, V38, P140
[7]   SAFETY OF PATIENTS WITH MEDICAL DEVICES DURING APPLICATION OF MAGNETIC-RESONANCE METHODS [J].
POHOST, GM ;
BLACKWELL, GG ;
SHELLOCK, FG .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES-SERIES, 1992, 649 :302-312
[8]   Vagus nerve stimulation (VNS) for treatment-resistant depressions: A multicenter study [J].
Rush, AJ ;
George, MS ;
Sackeim, HA ;
Marangell, LB ;
Husain, MM ;
Giller, C ;
Nahas, Z ;
Haines, S ;
Simpson, RK ;
Goodman, R .
BIOLOGICAL PSYCHIATRY, 2000, 47 (04) :276-286
[9]  
Shellock F.G., 1998, Pocket Guide to MR Procedures and Metallic Objects: Update 1998
[10]   Magnetic resonance imaging with implanted neurostimulators:: An in vitro and in vivo study [J].
Tronnier, VM ;
Staubert, A ;
Hähnel, S ;
Sarem-Aslani, A .
NEUROSURGERY, 1999, 44 (01) :118-125