Tin-filter Enhanced Dual-Energy-CT: Image Quality and Accuracy of CT Numbers in Virtual Noncontrast Imaging

被引:63
作者
Kaufmann, Sascha [1 ]
Sauter, Alexander [1 ]
Spira, Daniel [1 ]
Gatidis, Sergios [1 ]
Ketelsen, Dominik [1 ]
Heuschmid, Martin [1 ]
Claussen, Claus D. [1 ]
Thomas, Christoph [1 ]
机构
[1] Dept Radiol Diagnost & Intervent Radiol, D-72076 Tubingen, Germany
关键词
Dual energy; dual source; computed tomography; virtual noncontrast imaging; CLINICAL-APPLICATIONS; COMPUTED-TOMOGRAPHY; MULTIDETECTOR CT; DIFFERENTIATION; DISCRIMINATION; PERFORMANCE; SYSTEM; MASSES; AORTA;
D O I
10.1016/j.acra.2013.01.010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Objectives: To measure and compare the objective image quality of true noncontrast (TNC) images with virtual noncontrast (VNC) images acquired by tin-filter-enhanced, dual-source, dual-energy computed tomography (DECT) of upper abdomen. Materials and Methods: Sixty-three patients received unenhanced abdominal CT and enhanced abdominal DECT (100/140 kV with tin filter) in portal-venous phase. VNC images were calculated from the DECT datasets using commercially available software. The mean attenuation of relevant tissues and image quality were compared between the TNC and VNC images. Image quality was rated objectively by measuring image noise and the sharpness of object edges using-custom-designed software. Measurements were compared using Student two-tailed t-test. Correlation coefficients for tissue attenuation measurements between TNC and VNC were calculated and the relative deviations were illustrated using Bland-Altman plots. Results: Mean attenuation differences between TNC and VNC (HUTNC - HUVNC) image sets were as follows: right liver lobe -4.94 Hounsfield units (HU), left liver lobe -3.29 HU, vena cava -2.19 HU, spleen -7.46 HU, pancreas 1.29 HU, fat -11.14 HU, aorta 1.29 HU, bone marrow 36.83 HU (all P < .05); right kidney 0.46 HU, left kidney 0.56 HU, vena portae -0.48 HU and muscle -0.62 HU (nonsignificant). Good correlations between VNC and TNC series were observed for liver, vena portae, kidneys, pancreas, muscle and bone marrow (Pearson's correlation coefficient >= 0.75). Mean image noise was significantly higher in TNC images (P < .0001). Measurements of edge sharpness revealed no significant differences between VNC and TNC images (P = .19). Conclusion: The Hounsfield units in VNC images closely resemble TNC images in the majority of the organs of the upper abdomen (kidneys, liver, pancreas). In spleen and fat, Hounsfield numbers in VNC images are tend to be higher than in TNC images. VNC images show a low image noise and satisfactory edge sharpness. Other criteria of image quality and the depiction of certain lesions need to be evaluated additionally.
引用
收藏
页码:596 / 603
页数:8
相关论文
共 26 条
[1]
Dual-energy-CT of hypervascular liver lesions in patients with HCC: investigation of image quality and sensitivity [J].
Altenbernd, Jens ;
Heusner, Till A. ;
Ringelstein, Adrian ;
Ladd, Susanne C. ;
Forsting, Michael ;
Antoch, Gerald .
EUROPEAN RADIOLOGY, 2011, 21 (04) :738-743
[2]
CLINICAL APPLICATION OF COMPTON AND PHOTOELECTRIC RECONSTRUCTION IN COMPUTED TOMOGRAPHY - PRELIMINARY-RESULTS [J].
AVRIN, DE ;
MACOVSKI, A ;
ZATZ, LM .
INVESTIGATIVE RADIOLOGY, 1978, 13 (03) :217-222
[3]
Virtual unenhanced second generation dual-source CT of the liver: Is it time to discard the conventional unenhanced phase? [J].
Barrett, T. ;
Bowden, D. J. ;
Shaida, N. ;
Godfrey, E. M. ;
Taylor, A. ;
Lomas, D. J. ;
Shaw, A. S. .
EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (07) :1438-1445
[4]
Abdominal Aorta: Evaluation with Dual-Source Dual-Energy Multidetector CT after Endovascular Repair of Aneurysms-Initial Observations [J].
Chandarana, Hersh ;
Godoy, Myrna C. B. ;
Vlahos, Ioannis ;
Graser, Anno ;
Babb, James ;
Leidecker, Christianne ;
Macari, Michael .
RADIOLOGY, 2008, 249 (02) :692-700
[5]
Estimating Effective Dose for CT Using Dose-Length Product Compared With Using Organ Doses: Consequences of Adopting International Commission on Radiological Protection Publication 103 or Dual-Energy Scanning [J].
Christner, Jodie A. ;
Kofler, James M. ;
McCollough, Cynthia H. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (04) :881-889
[6]
Dual energy CT (DECT) of the liver: conventional versus virtual unenhanced images [J].
De Cecco, Carlo Nicola ;
Buffa, Vitaliano ;
Fedeli, Stefano ;
Luzietti, Monica ;
Vallone, Andrea ;
Ruopoli, Roberto ;
Miele, Vittorio ;
Rengo, Marco ;
Paolantonio, Pasquale ;
Enrici, Michelangelo Maurizi ;
Laghi, Andrea ;
David, Vincenzo .
EUROPEAN RADIOLOGY, 2010, 20 (12) :2870-2875
[7]
First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[8]
Dual-Energy CT in Patients Suspected of Having Renal Masses: Can Virtual Nonenhanced Images Replace True Nonenhanced Images? [J].
Graser, Anno ;
Johnson, Thorsten R. C. ;
Hecht, Elizabeth M. ;
Becker, Christoph R. ;
Leidecker, Christianne ;
Staehler, Michael ;
Stief, Christian G. ;
Hildebrandt, Henriette ;
Godoy, Myrna C. B. ;
Finn, Myra E. ;
Stepansky, Flora ;
Reiser, Maximilian F. ;
Macari, Michael .
RADIOLOGY, 2009, 252 (02) :433-440
[9]
Dual energy CT: preliminary observations and potential clinical applications in the abdomen [J].
Graser, Anno ;
Johnson, Thorsten R. C. ;
Chandarana, Hersh ;
Macari, Michael .
EUROPEAN RADIOLOGY, 2009, 19 (01) :13-23
[10]
COMPUTERIZED TRANSVERSE AXIAL SCANNING (TOMOGRAPHY) .1. DESCRIPTION OF SYSTEM [J].
HOUNDFIELD, GN .
BRITISH JOURNAL OF RADIOLOGY, 1973, 46 (552) :1016-1022