Spectral surface plasmon resonance biosensor for detection of staphylococcal enterotoxin B in milk

被引:265
作者
Homola, J
Dostálek, J
Chen, SF
Rasooly, A
Jiang, SY
Yee, SS
机构
[1] Univ Washington, Dept Elect Engn, Seattle, WA 98195 USA
[2] Univ Washington, Dept Chem Engn, Seattle, WA 98195 USA
[3] US FDA, Div Microbiol Studies, Washington, DC 20204 USA
基金
美国国家科学基金会;
关键词
biosensor; surface plasmon resonance; staphylococcal enterotoxin B;
D O I
10.1016/S0168-1605(02)00010-7
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
This work evaluates a newly developed wavelength modulation-based SPR biosensor for the detection of staphylococcal enterotoxin B (SEB) in milk. Two modes of operation of the SPR biosensor are described: direct detection of SEB and sandwich assay. In the sandwich assay detection mode, secondary antibodies are bound to the already captured Toxin to amplify sensor response. Samples including SEB in buffer and SEB in milk were analyzed in this work. The SPR biosensor has been shown to be capable of directly detecting concentrations of SEB in buffer as low as 5 ng/ml. In sandwich detection mode, the lowest detection limit was determined to be 0.5 ng/ml for both buffer and milk samples. The reported wavelength modulation-based SPR sensor provides a generic platform which can be tailored for detection of various foodborne pathogens and agents for food analysis and testing. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:61 / 69
页数:9
相关论文
共 28 条
[1]  
ARCHER D L, 1988, Clinical Microbiology Reviews, V1, P377
[2]  
BERGDOLL MS, 1991, J ASSOC OFF ANA CHEM, V74, P706
[3]  
BERGDOLL MS, 1995, ZBL BAKT-INT J MED M, V282, P1
[4]  
Boisde G, 1996, CHEM BIOCH SENSING O
[5]   BIOCHEMICAL ASPECTS OF BIOSENSORS [J].
BYFIELD, MP ;
ABUKNESHA, RA .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (4-5) :373-400
[6]   Piezoelectric immunosensor for the detection of immunoglobulin M [J].
Chu, X ;
Lin, ZH ;
Shen, GL ;
Yu, RQ .
ANALYST, 1995, 120 (12) :2829-2832
[7]   Rapid detection of Escherichia coli O157:H7 in ground beef using a fiber-optic biosensor [J].
DeMarco, DR ;
Saaski, EW ;
McCrae, DA ;
Lim, DV .
JOURNAL OF FOOD PROTECTION, 1999, 62 (07) :711-716
[8]   ESTIMATION OF HUMAN DOSE OF STAPHYLOCOCCAL ENTEROTOXIN-A FROM A LARGE OUTBREAK OF STAPHYLOCOCCAL FOOD POISONING INVOLVING CHOCOLATE MILK [J].
EVENSON, ML ;
HINDS, MW ;
BERNSTEIN, RS ;
BERGDOLL, MS .
INTERNATIONAL JOURNAL OF FOOD MICROBIOLOGY, 1988, 7 (04) :311-316
[9]   Detection of Escherichia coli O157:H7 using a surface plasmon resonance biosensor [J].
Fratamico, PM ;
Strobaugh, TP ;
Medina, MB ;
Gehring, AG .
BIOTECHNOLOGY TECHNIQUES, 1998, 12 (07) :571-576
[10]  
GAUGLITZ G, 1996, OPTOCHEMICAL OPTOIMM, V1