Overestimation of stent delivery balloon diameters by manufacturers' compliance tables: A quantitative coronary analysis of duet and NIR stent implantation

被引:13
作者
Hehrlein, C
DeVries, JJ
Wood, TA
Haller, SD
Arab, A
Kloostra, AC
Lauer, MA
Fischell, TA
机构
[1] Univ Freiburg, Dept Cardiol, D-79106 Freiburg, Germany
[2] Borgess Med Ctr, Inst Heart, Kalamazoo, MI USA
关键词
angioplasty; coronary angiography; stents;
D O I
10.1002/ccd.1205
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Although manufacturers' compliance tables of stent delivery balloons indicate the diameter of the balloon at a given inflation pressure, it is unclear whether these data correlate with in vivo true intracoronary balloon diameters (TBDs). The TBDs of two new-generation balloon-expandable stent delivery systems (Duet and NIR) were measured by quantitative coronary analysis (QCA) in 100 consecutive patients. The manufacturers' stated balloon diameter (BD) of the stent delivery systems overestimated the TBD in 94% +/- 4% of patients receiving both Duet or NIR stent implantations. In only 6% of the patients, the TBD matched the manufacturers' stated balloon diameter. There was no underestimation of TBDs by both manufacturers' compliance tables. The Duet tables overestimated TBDs by 14% +/- 8% (range, 1%-36%), the NIR tables by 18% +/- 8% (range, 1%-41%), P < 0.05, Duet vs. NIR, respectively. When the manufacturers' data were corrected for the differences in reporting data from in vitro tests, i.e., balloon compliance data with or without the stent, the degree of overestimation of diameters was similar for Duet and NIR stem delivery balloons (14% +/- 8% vs. 13% +/- 7%, Duet vs. NIR; P = NS). Manufacturers' compliance tables of both the Duet and NIR stent delivery balloon systems significantly overestimate the true intracoronary balloon diameter. The manufacturers' of stent delivery balloons should clearly state on the box, if balloon compliance data were derived from in vitro bench testing, which phantoms were used for compliance analysis, and that the tables may overestimate the true intracoronary balloon diameter. The findings of the present study have important clinical implications with respect to performing coronary stent implantation with precision. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:474 / 478
页数:5
相关论文
共 10 条
[1]  
Baim DS, 1998, CIRCULATION, V98, P661
[2]   Mechanisms of residual lumen stenosis after high-pressure stent implantation -: A quantitative coronary angiography and intravascular ultrasound study [J].
Bermejo, J ;
Botas, J ;
García, E ;
Elízaga, J ;
Osende, J ;
Soriano, J ;
Abeytua, M ;
Delcán, JL .
CIRCULATION, 1998, 98 (02) :112-118
[3]  
CRITTENDEN JF, 1987, Z KARDIOL, V76, P33
[4]   Procedural and follow up results with a new balloon expandable stent in unselected lesions [J].
di Mario, C ;
Reimers, B ;
Almagor, Y ;
Moussa, I ;
Di Francesco, L ;
Ferraro, M ;
Leon, MB ;
Richter, K ;
Colombo, A .
HEART, 1998, 79 (03) :234-241
[5]   Rupture of PTCA balloons - an in vitro examination [J].
Dubel, HP ;
Gliech, V ;
Bunk, G ;
Kubat, B ;
Schneider, M ;
Kleber, FX .
ZEITSCHRIFT FUR KARDIOLOGIE, 1997, 86 (12) :968-974
[6]   PERMANENT INFLATION OF DETACHABLE BALLOONS WITH A LOW-VISCOSITY, HYDROPHILIC POLYMERIZING SYSTEM [J].
GOTO, K ;
HALBACH, VV ;
HARDIN, CW ;
HIGASHIDA, RT ;
HIESHIMA, GB .
RADIOLOGY, 1988, 169 (03) :787-790
[7]   ANGIOGRAPHIC PATTERNS OF BALLOON INFLATION DURING PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY - ROLE OF PRESSURE-DIAMETER CURVES IN STUDYING DISTENSIBILITY AND ELASTICITY OF THE STENOTIC LESION AND THE MECHANISM OF DILATION [J].
HJEMDAHLMONSEN, CE ;
AMBROSE, JA ;
BORRICO, S ;
COHEN, M ;
SHERMAN, W ;
ALEXOPOULOS, D ;
GORLIN, R ;
FUSTER, V .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :569-575
[8]   EFFECT OF INFLATION PRESSURES ON CORONARY ANGIOPLASTY BALLOONS [J].
JAIN, A ;
DEMER, LL ;
RAIZNER, AE ;
ROBERTS, R .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (01) :26-28
[9]  
Kereiakes Dean J., 1999, Journal of the American College of Cardiology, V33, p95A
[10]   Angioplasty balloon compliance: Can in vivo size be predicted from in vitro pressure profile measurements? [J].
More, RS ;
Brack, MJ ;
Gershlick, AH .
CLINICAL CARDIOLOGY, 1996, 19 (05) :393-397