Angioplasty balloon compliance: Can in vivo size be predicted from in vitro pressure profile measurements?

被引:5
作者
More, RS [1 ]
Brack, MJ [1 ]
Gershlick, AH [1 ]
机构
[1] GLENFIELD GEN HOSP,ACAD DEPT CARDIOL,LEICESTER LE3 9QP,LEICS,ENGLAND
关键词
balloon size; intracoronary; pressure profile; quantitative angiography; in vitro measurements;
D O I
10.1002/clc.4960190512
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and hypothesis: This study was undertaken to determine whether the behavior of angioplasty balloons within coronary arteries may differ from that anticipated from data provided by the manufacturers. In particular, the in vitro pressure-diameter profiles may not truly represent in vivo sizes. Methods: Thus, we assessed the degree of correlation of in vitro with in vivo measurements obtained during routine angioplasty practice. In vivo size of 2.5 mm compliant (n = 8) and 3 mm semicompliant (n = 8) balloons was assessed using quantitative angiography for first, second, and third inflations Results: In vivo size was less than expected from in vitro measurements. In general balloon diameter increased with inflation pressures up to 8 atmospheres, and some degree of elastic recoil was evident with both balloon types after the last inflation. Conclusion: In vivo balloon size may not be accurately predicted from manufacturers' published data. Size is more likely to be affected by factors such as lesion characteristics and elasticity of the vessel wall than by balloon material compliance characteristics.
引用
收藏
页码:393 / 397
页数:5
相关论文
共 13 条
[1]   EVALUATION OF A PERSONAL COMPUTER-BASED QUANTITATIVE CORONARY ANALYSIS SYSTEM FOR RAPID ASSESSMENT OF CORONARY STENOSES [J].
ALVAREZ, LG ;
JACKSON, SA ;
BERRY, JA ;
EICHHORN, EJ .
AMERICAN HEART JOURNAL, 1992, 123 (06) :1500-1510
[2]   CORONARY-ARTERY DILATATION REQUIRING HIGH INFLATION PRESSURE [J].
BUSH, CA ;
RYAN, JM ;
ORSINI, AR ;
HENNEMANN, WW .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1991, 22 (02) :112-114
[3]  
GRAY WA, 1991, CIRCULATION S2, V2, P591
[4]   INFLUENCE OF BALLOON SIZE AND STENOSIS MORPHOLOGY ON IMMEDIATE AND DELAYED ELASTIC RECOIL AFTER PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY [J].
HANET, C ;
WIJNS, W ;
MICHEL, X ;
SCHROEDER, E .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (02) :506-511
[5]   METHODOLOGICAL PROBLEMS RELATED TO THE QUANTITATIVE ASSESSMENT OF STRETCH, ELASTIC RECOIL, AND BALLOON-ARTERY RATIO [J].
HERMANS, WRM ;
RENSING, BJ ;
STRAUSS, BH ;
SERRUYS, PW .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1992, 25 (03) :174-185
[6]   ANGIOGRAPHIC PATTERNS OF BALLOON INFLATION DURING PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY - ROLE OF PRESSURE-DIAMETER CURVES IN STUDYING DISTENSIBILITY AND ELASTICITY OF THE STENOTIC LESION AND THE MECHANISM OF DILATION [J].
HJEMDAHLMONSEN, CE ;
AMBROSE, JA ;
BORRICO, S ;
COHEN, M ;
SHERMAN, W ;
ALEXOPOULOS, D ;
GORLIN, R ;
FUSTER, V .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :569-575
[7]   INFLATION PRESSURE REQUIREMENTS DURING CORONARY ANGIOPLASTY [J].
KAHN, JK ;
RUTHERFORD, BD ;
MCCONAHAY, DR ;
HARTZLER, GO .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 21 (03) :144-147
[8]  
LEHMANN KG, 1989, CIRCULATION S2, V80, P372
[9]  
MONSON CE, 1990, J AM COLL CARDIOL, V16, P569
[10]   EFFECT OF BALLOON MATERIAL ON CORONARY ANGIOPLASTY [J].
MOONEY, MR ;
MOONEY, JF ;
LONGE, TF ;
BRANDENBURG, RO .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (17) :1481-1482