Qualitative and Quantitative Assessment of Metal Artifacts Arising from Implantable Cardiac Pacing Devices in Oncological PET/CT Studies: A Phantom Study

被引:13
作者
Ay, Mohammad R. [1 ,2 ,3 ,6 ]
Mehranian, Abolfazl [1 ,2 ]
Abdoli, Mehrsima [4 ]
Ghafarian, Pardis [2 ,5 ]
Zaidi, Habib [4 ,6 ,7 ]
机构
[1] Univ Tehran Med Sci, Dept Med Phys & Biomed Engn, Tehran, Iran
[2] Univ Tehran Med Sci, Res Ctr Sci & Technol Med, Tehran, Iran
[3] Univ Tehran Med Sci, Res Inst Nucl Med, Tehran, Iran
[4] Univ Groningen, Univ Med Ctr Groningen, Dept Nucl Med & Mol Imaging, Groningen, Netherlands
[5] Shahid Beheshti Univ, Dept Radiat Med, Tehran, Iran
[6] Univ Hosp, Div Nucl Med Geneva, CH-1211 Geneva, Switzerland
[7] Univ Geneva, Geneva Neurosci Ctr, CH-1211 Geneva, Switzerland
基金
瑞士国家科学基金会;
关键词
PET/CT; Attenuation correction; Metal artifact reduction; Pacemaker; Defibrillator; Misalignment; ATTENUATION CORRECTION; EMISSION TOMOGRAPHY; DENTAL IMPLANTS; NECK; HEAD; REDUCTION; IMPACT; MAP;
D O I
10.1007/s11307-010-0467-x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: We evaluate the magnitude of metallic artifacts caused by various implantable cardiac pacing devices (without leads) on both attenuation maps (mu-maps) and positron emission tomography (PET) images using experimental phantom studies. We also assess the efficacy of a metal artifact reduction (MAR) algorithm along with the severity of artifacts in the presence of misalignment between mu-maps and PET images. Methods: Four pacing devices including two pacemakers (pacemakers 1 and 2) and two cardiac resynchronization therapy (CRT) devices of pacemaker (CRT-P) and defibrillator (CRT-D) type were placed in three phantoms including a cylindrical Ge-68 phantom, a water-bath phantom and an anthropomorphic heart/thorax phantom. The mu-maps were derived from computed tomography (CT) images reconstructed using the standard method supplied by the manufacturer and those reconstructed using the MAR algorithm. In addition, the standard reconstructed CT images of the last two phantoms were manually misaligned by 10 mm along the patient's axis to simulate misalignment between CT and PET images. Results: The least and severest artifacts produced on both mu-maps and PET images of the Ge-68 phantom were induced by CRT-P and pacemaker 1 devices, respectively. In the water-bath phantom, CRT-P induced 17.5% over- and 9.2% underestimation of tracer uptake whereas pacemaker 1 induced 69.6% over- and 65.7% underestimation. In the heart/thorax phantom representing a pacemaker-bearing patient, pacemaker 1 induced 41.8% increase and 36.6% decrease in tracer uptake and attenuation coefficients on average in regions corresponding to bright and dark streak artifacts, respectively. Statistical analysis revealed that the MAR algorithm was successful in reducing bright streak artifacts, yet unsuccessful for dark ones. In the heart/thorax phantom, the MAR algorithm reduced the overestimations to 4.4% and the underestimations to 35.5% on average. Misalignment between mu-maps and PET images increased the peak of pseudo-uptake by approximately 20%. Conclusions: This study demonstrated that, depending on their elemental composition, different implantable cardiac pacing devices result in varying magnitudes of metal artifacts and thus pseudo-uptake on PET images. The MAR algorithm was not successful in compensating for underestimations which calls for a more efficient algorithm. The results showed that misalignments between PET and CT images render metal-related pseudo-uptake more severe.
引用
收藏
页码:1077 / 1087
页数:11
相关论文
共 22 条
[1]
Reduction of dental filling metallic artifacts in CT-based attenuation correction of PET data using weighted virtual sinograms optimized by a genetic algorithm [J].
Abdoli, Mehrsima ;
Ay, Mohammad Reza ;
Ahmadian, Alireza ;
Dierckx, Rudi A. J. O. ;
Zaidi, Habib .
MEDICAL PHYSICS, 2010, 37 (12) :6166-6177
[2]
A virtual sinogram method to reduce dental metallic implant artefacts in computed tomography-based attenuation correction for PET [J].
Abdoli, Mehrsima ;
Ay, Mohammad Reza ;
Ahmadian, Alireza ;
Zaidi, Habib .
NUCLEAR MEDICINE COMMUNICATIONS, 2010, 31 (01) :22-31
[3]
Attenuation-emission alignment in cardiac PET/CT based on consistency conditions [J].
Alessio, Adam M. ;
Kinahan, Paul E. ;
Champley, Kyle M. ;
Caldwell, James H. .
MEDICAL PHYSICS, 2010, 37 (03) :1191-1200
[4]
Transmission scanning in emission tomography [J].
Bailey, DL .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (07) :774-787
[5]
Bujenovic Steve, 2003, Mol Imaging Biol, V5, P20, DOI 10.1016/S1536-1632(03)00037-4
[6]
DiFilippo FP, 2005, J NUCL MED, V46, P436
[7]
Artifacts at PET and PET/CT caused by metallic hip prosthetic material [J].
Goerres, GW ;
Ziegler, SI ;
Burger, C ;
von Schulthess, TBGK ;
Buck, A .
RADIOLOGY, 2003, 226 (02) :577-584
[8]
Head and neck imaging with PET and PET/CT: artefacts from dental metallic implants [J].
Goerres, GW ;
Hany, TF ;
Kamel, E ;
von Schulthess, GK ;
Buck, A .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (03) :367-370
[9]
Halpern BS, 2004, J NUCL MED, V45, P290
[10]
A knowledge-based method for reducing attenuation artefacts caused by cardiac appliances in myocardial PET/CT [J].
Hamill, James J. ;
Brunken, Richard C. ;
Bybel, Bohdan ;
DiFilippo, Frank P. ;
Faul, David D. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (11) :2901-2918