Artifacts at PET and PET/CT caused by metallic hip prosthetic material

被引:133
作者
Goerres, GW
Ziegler, SI
Burger, C
von Schulthess, TBGK
Buck, A
机构
[1] Univ Zurich Hosp, Div Nucl Med, CH-8091 Zurich, Switzerland
[2] Tech Univ Munich, Nucl Med Clin, D-8000 Munich, Germany
关键词
images; processing; phantoms; positron emission tomography (PET); artifact; stents and prostheses; test objects;
D O I
10.1148/radiol.2262012141
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Hip prosthetic material and a steel rod were scanned in a water bath of fluorine 18 fluorodeoxyglucose (FDG) with positron emission tomographic (PET) and PET/computed tomographic (CT) scanners to evaluate the generation of artifacts adjacent to the metal. The influences of attenuation correction (AC), positioning of the object, and image reconstruction were examined. Use of CT- and germanium 68- based AC resulted in generation of artifacts that mimicked increased FDG uptake. These artifacts were more evident when the object was moved between the emission and transmission scans. When attenuation-weighted iterative reconstruction was used, these artifacts were less evident. (C) RSNA, 2002.
引用
收藏
页码:577 / 584
页数:8
相关论文
共 13 条
[1]
An automatic classification technique for attenuation correction in positron emission tomography [J].
Bettinardi, V ;
Pagani, E ;
Gilardi, MC ;
Landoni, C ;
Riddell, C ;
Rizzo, G ;
Castiglioni, I ;
Belluzzo, D ;
Lucignani, G ;
Schubert, S ;
Fazio, F .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (05) :447-458
[2]
Fast reconstruction of 3D PET data with accurate statistical modeling [J].
Comtat, C ;
Kinahan, PE ;
Defrise, M ;
Michel, C ;
Townsend, DW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :1083-1089
[3]
Chronic osteomyelitis: Detection with FDG PET and correlation with histopathologic findings [J].
Guhlmann, A ;
Brecht-Krauss, D ;
Sugar, G ;
Glatting, G ;
Kotzerke, J ;
Kinzl, L ;
Reske, SN .
RADIOLOGY, 1998, 206 (03) :749-754
[4]
Heiba S I., 2000, Clin Positron Imaging, V3, P237, DOI 10.1016/S1095-0397(01)00054-1
[5]
Fluorine-18 fluorodeoxyglucose PET in infectious bone diseases:: results of histologically confirmed cases [J].
Kälicke, T ;
Schmitz, A ;
Risse, JH ;
Arens, S ;
Keller, E ;
Hansis, M ;
Schmitt, O ;
Biersack, HJ ;
Grünwald, F .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (05) :524-528
[6]
Attenuation correction for a combined 3D PET/CT scanner [J].
Kinahan, PE ;
Townsend, DW ;
Beyer, T ;
Sashin, D .
MEDICAL PHYSICS, 1998, 25 (10) :2046-2053
[7]
A JAVA']JAVA environment for medical image data analysis: initial application for brain PET quantitation [J].
Mikolajczyk, K ;
Szabatin, M ;
Rudnicki, P ;
Grodzki, M ;
Burger, C .
MEDICAL INFORMATICS, 1998, 23 (03) :207-214
[8]
Chronic osteomyelitis of the femur: value of PET imaging [J].
Robiller, FC ;
Stumpe, KDM ;
Kossmann, T ;
Weisshaupt, D ;
Bruder, E ;
von Schulthess, GK .
EUROPEAN RADIOLOGY, 2000, 10 (05) :855-858
[9]
Seabold JE, 1999, Q J NUCL MED, V43, P21
[10]
Infection imaging using whole-body FDG-PET [J].
Stumpe, KDM ;
Dazzi, H ;
Schaffner, A ;
von Schulthess, GK .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (07) :822-832