The determination of trace metals in estuarine and coastal waters using a voltammetric in situ profiling system

被引:25
作者
Howell, KA
Achterberg, EP [1 ]
Braungardt, CB
Tappin, AD
Turner, DR
Worsfold, PJ
机构
[1] Univ Plymouth, Sch Environm Sci, Plymouth Environm Res Ctr, Plymouth PL4 8AA, Devon, England
[2] Univ Gothenburg, Dept Analyt & Marine Chem, S-341296 Gothenburg, Sweden
关键词
D O I
10.1039/b300712j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work presents the optimisation, validation and field deployment of a voltammetric in situ profiling (VIP) system for the simultaneous determinations of dynamic Cd(II), Cu(II) and Pb(II) in estuarine and coastal waters. Systematic studies in NaNO3 (as a supporting electrolyte) and seawater, indicated that variations in ionic strength, pH and dissolved oxygen did not affect the response of the instrument, whereas an Arrhenius type temperature response was observed. The VIP instrument allows the determination of 2-3 samples h(-1), and has a detection limit (defined as 3sigma) in seawater for Cd(II): 23 pM, Cu(II): 1.13 nM, and Pb(II): 23 pM. The VIP system accurately measured the total dissolved concentrations of Cd(II), Cu(II) and Pb(II) in two certified reference materials; SLRS-3, a river water, and SLEW-2, an estuarine water. Field evaluation of the instrumentation and analytical methods was achieved through a series of surveys in the Plym Estuary (Devon, UK), from which environmental data are presented.
引用
收藏
页码:734 / 741
页数:8
相关论文
共 29 条
[21]   REPRODUCIBLE, RELIABLE AND RUGGED HG-PLATED IR-BASED MICROELECTRODE FOR IN-SITU MEASUREMENTS IN NATURAL-WATERS [J].
TERCIER, ML ;
PARTHASARATHY, N ;
BUFFLE, J .
ELECTROANALYSIS, 1995, 7 (01) :55-63
[22]  
Tercier-Waeber ML, 2001, ABSTR PAP AM CHEM S, V221, pU583
[23]   Real-time continuous Mn(II) monitoring in lakes using a novel voltammetric in situ profiling system [J].
Tercier-Waeber, ML ;
Belmont-Hebert, C ;
Buffle, J .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1998, 32 (10) :1515-1521
[24]  
Tercier-Waeber ML, 1999, SEA TECHNOL, V40, P74
[25]   Submersible voltammetric probes for in situ real-time trace element measurements in surface water, groundwater and sediment-water interface [J].
Tercier-Waeber, ML ;
Buffle, J ;
Confalonieri, F ;
Riccardi, G ;
Sina, A ;
Graziottin, F ;
Fiaccabrino, GC ;
Koudelka-Hep, M .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1999, 10 (12) :1202-1213
[26]   Submersible online oxygen removal system coupled to an in situ voltammetric probe for trace element monitoring in freshwater [J].
Tercier-Waeber, ML ;
Buffle, J .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2000, 34 (18) :4018-4024
[28]   POTENTIALS AND POTENTIALITIES OF CATHODIC STRIPPING VOLTAMMETRY OF TRACE-ELEMENTS IN NATURAL-WATERS [J].
VAN DEN BERG, CMG .
ANALYTICA CHIMICA ACTA, 1991, 250 (01) :265-276
[29]   Renewable-reagent electrochemical sensor for monitoring trace metal contaminants [J].
Wang, J ;
Lu, JM ;
Luo, DB ;
Wang, JY ;
Jiang, M ;
Tian, BM ;
Olsen, K .
ANALYTICAL CHEMISTRY, 1997, 69 (14) :2640-2645